We have located links that may give you full text access.
Case Reports
Journal Article
Complete pacemaker failure following lightning strike injury: A case report.
Pacing and Clinical Electrophysiology : PACE 2024 August 19
INTRODUCTION: Modern cardiovascular implantable electronic devices (CIEDs) have mechanisms that prevent damage from external electric shocks, and malfunction following accidental electrocution is rare. However, the effects of lightning injuries in patients with CIEDs are uncertain.
CASE PRESENTATION: A 74-year-old man with a dual-chamber pacemaker due to complete heart block was struck by a lightning while farming. He had no serious injury at the time and sought medical evaluation 1 month later, when he presented with asymptomatic bradycardia. Device interrogation suggested major battery and lead damage, requiring extraction and subsequent placement of a new pacing system.
DISCUSSION: While a previous report depicted pacing threshold elevation without extensive device impairment, our patient presented with major damage to the whole pacing system. The factors contributing to these divergent outcomes are unclear. Differences in injury mechanism, pacemaker model, and the pattern of electric current dispersion within the device may each play a part in this discrepancy.
CASE PRESENTATION: A 74-year-old man with a dual-chamber pacemaker due to complete heart block was struck by a lightning while farming. He had no serious injury at the time and sought medical evaluation 1 month later, when he presented with asymptomatic bradycardia. Device interrogation suggested major battery and lead damage, requiring extraction and subsequent placement of a new pacing system.
DISCUSSION: While a previous report depicted pacing threshold elevation without extensive device impairment, our patient presented with major damage to the whole pacing system. The factors contributing to these divergent outcomes are unclear. Differences in injury mechanism, pacemaker model, and the pattern of electric current dispersion within the device may each play a part in this discrepancy.
Full text links
Related Resources
Trending Papers
Central Nervous System Involvement in Systemic Autoimmune Rheumatic Diseases-Diagnosis and Treatment.Pharmaceuticals 2024 August 7
Sedation for awake tracheal intubation: A systematic review and network meta-analysis.Anaesthesia 2024 October 28
Efficacy of Traditional Anti-lipidemic Drugs in Lowering Lipoprotein(a) Levels: A Systematic Review.Curēus 2024 September
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.