Red filter meibography by smartphones in patients with meibomian gland dysfunction: a validity and reliability study.
BMJ Open Ophthalmology 2024 April 13
OBJECTIVE: The objective of this study is to determine the validity and reliability of the red filter meibography by smartphone compared with infrared in assessing meibomian gland drop-out.
METHODS AND ANALYSIS: An analytical cross-sectional study was done with a total of 35 subjects (68 eyes) with suspected MGD based on symptoms and lid morphological abnormalities. Meibomian glands were photographed using two smartphones (Samsung S9 and iPhone XR) on a slit-lamp with added red filter. Images were assessed subjectively using meiboscore by the two raters and drop-out percentages were assessed by ImageJ.
RESULTS: There was no agreement in meiboscore and a minimal level of agreement in drop-out percentages between red filter meibography and infrared. Inter-rater reliability showed no agreement between two raters. Intra-rater reliability demonstrated weak agreement in rater 1 and no agreement in rater 2.
CONCLUSION: Validity of the red filter meibography technique by smartphones is not yet satisfactory in evaluating drop-out. Further improvement on qualities of images must be done and research on subjective assessment was deemed necessary due to poor results of intrarater and inter-rater reliability.
METHODS AND ANALYSIS: An analytical cross-sectional study was done with a total of 35 subjects (68 eyes) with suspected MGD based on symptoms and lid morphological abnormalities. Meibomian glands were photographed using two smartphones (Samsung S9 and iPhone XR) on a slit-lamp with added red filter. Images were assessed subjectively using meiboscore by the two raters and drop-out percentages were assessed by ImageJ.
RESULTS: There was no agreement in meiboscore and a minimal level of agreement in drop-out percentages between red filter meibography and infrared. Inter-rater reliability showed no agreement between two raters. Intra-rater reliability demonstrated weak agreement in rater 1 and no agreement in rater 2.
CONCLUSION: Validity of the red filter meibography technique by smartphones is not yet satisfactory in evaluating drop-out. Further improvement on qualities of images must be done and research on subjective assessment was deemed necessary due to poor results of intrarater and inter-rater reliability.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app