Reducing umbilical catheter migration rates by using a novel securement device.

OBJECTIVE: This study evaluates the effectiveness of a novel device, LifeBubble, in reducing umbilical cord catheter (UC) migration and associated complications in neonates.

STUDY DESIGN: A retrospective review was performed at Oregon Health & Science University's NICU (2019-2021) to compare standard adhesive securement with LifeBubble. The primary outcomes were UC migration, discontinuation due to malposition, and CLABSI incidence. Differences between groups were statistically analyzed and logistic regression used to adjust for potential confounders.

RESULTS: Among 118 neonates (57 LifeBubble, 61 adhesive), LifeBubble significantly reduced migration of any UC > 1 vertebral body (12.3% vs. 55.7%), including UVC migration (5.3% vs. 39.3%) and UAC migration (7.0% vs 23.0%), as well as UVC discontinuation due to malposition (5.6% vs 37.7%). The number needed to treat (NNT) to prevent one instance of UVC discontinuation is 4.

CONCLUSION: LifeBubble effectively reduces UC migration and premature discontinuation, indicating its potential to enhance neonatal care and safety.

Full text links

We have located open access text paper links.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app