We have located links that may give you full text access.
Utilization of the Reinstein ICL Sizing Formula With Hand-held Ultrasound Biomicroscopy Measurements.
Journal of Refractive Surgery 2024 March
PURPOSE: To evaluate the accuracy of the Reinstein formula with hand-held ultrasound biomicroscopy (UBM) measurements for sizing of the Implantable Collamer Lens (ICL).
METHODS: A total of 107 myopic eyes of 57 patients implanted with the ICL were included in the study. The size of the ICL was selected based on the manufacturer's recommendations. Agreement between the vault predicted by the Reinstein formula and the vault measured postoperatively was analyzed with Bland-Altman plots.
RESULTS: A total of 95% and 81% of patients had a postoperative vault ranging from 150 to 1,000 and 250 to 750 μm, respectively. The mean vault predicted by the Reinstein formula and the postoperative vault in the current study were 580 ± 181 and 547 ± 200 μm, respectively. The size recommendations of the Reinstein formula and the formula provided by the manufacturer, the Kojima formula, and the Dougherty formula overlapped in 50%, 57%, and 49% of eyes, respectively.
CONCLUSIONS: The results show that the Reinstein formula combined with a hand-held UBM provides reliable sizing predictions of the ICL. However, considering that robotic UBM measurements have demonstrated a narrower range of deviation in predicting vault depth in previous studies, a direct comparison study between robotic UBM and hand-held UBM measurements is necessary to fully assess the limitations of combining hand-held UBM with the Reinstein formula. [ J Refract Surg . 2024;40(3):e142-e147.] .
METHODS: A total of 107 myopic eyes of 57 patients implanted with the ICL were included in the study. The size of the ICL was selected based on the manufacturer's recommendations. Agreement between the vault predicted by the Reinstein formula and the vault measured postoperatively was analyzed with Bland-Altman plots.
RESULTS: A total of 95% and 81% of patients had a postoperative vault ranging from 150 to 1,000 and 250 to 750 μm, respectively. The mean vault predicted by the Reinstein formula and the postoperative vault in the current study were 580 ± 181 and 547 ± 200 μm, respectively. The size recommendations of the Reinstein formula and the formula provided by the manufacturer, the Kojima formula, and the Dougherty formula overlapped in 50%, 57%, and 49% of eyes, respectively.
CONCLUSIONS: The results show that the Reinstein formula combined with a hand-held UBM provides reliable sizing predictions of the ICL. However, considering that robotic UBM measurements have demonstrated a narrower range of deviation in predicting vault depth in previous studies, a direct comparison study between robotic UBM and hand-held UBM measurements is necessary to fully assess the limitations of combining hand-held UBM with the Reinstein formula. [ J Refract Surg . 2024;40(3):e142-e147.] .
Full text links
Related Resources
Trending Papers
Updated evidence on cardiovascular and renal effects of GLP-1 receptor agonists and combination therapy with SGLT2 inhibitors and finerenone: a narrative review and perspectives.Cardiovascular Diabetology 2024 November 15
Methods for determining optimal positive end-expiratory pressure in patients undergoing invasive mechanical ventilation: a scoping review.Canadian Journal of Anaesthesia 2024 November 20
Cardiac Failure and Cardiogenic Shock: Insights Into Pathophysiology, Classification, and Hemodynamic Assessment.Curēus 2024 October
The Management of Interstitial Lung Disease in the ICU: A Comprehensive Review.Journal of Clinical Medicine 2024 November 6
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app