We have located links that may give you full text access.
CG02N Analyzer Accurately Measures Fibrinogen Level in Whole Blood in the Presence of Low- or High-Dose Heparin.
Indian Journal of Hematology & Blood Transfusion 2024 January
Rapid fibrinogen (Fbg) evaluation is important in patients with massive bleeding during severe trauma and those undergoing major surgery. However, there are only a few studies on the point-of-care Fbg analyzer. In this study, we aimed to investigate the accuracy of Fbg level measured using CG02N, with whole blood contained in lithium-heparinized syringes with two different concentrations of heparin. Blood samples were collected in lithium-heparinized tubes, namely PREZA-PAK®II (low-dose heparin group [LG], 7 IU/mL) and Pro-Vent® Plus (high-dose heparin group [HG], 70.5 IU/mL). The Fbg levels in LG and HG were compared with those of citrated plasma Fbg (standard-Fbg). Strong correlations with respect to the Fbg level were observed between standard-Fbg and LG or HG (r = 0.968, p < 0.0001; r = 0.970, p < 0.0001, respectively). We demonstrated that the Fbg level in whole-blood samples was accurately assessed by CG02N and not affected by low- or high-dose heparin.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2025 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app