Validation of the automated oscillometric upper-arm cuff home blood pressure monitor AVITA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.
Blood Pressure Monitoring 2023 November 6
OBJECTIVE: This study evaluated the accuracy of the automated oscillometric upper-arm cuff device AVITA BPM82 intended for home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01.
METHODS: Participants were recruited to fulfill the age, sex, blood pressure (BP), and limb distribution criteria of the AAMI/ESH/ISO Universal Standard and its Amendment 1.2020-01 in a general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-33 (medium) and 33-42 cm (large).
RESULTS: One-hundred-one individuals were recruited and 85 were analyzed [mean age 57.3 ± 15.1 (SD) years, 46 men, arm circumference 32 ± 5.1 cm, range 22.3-42 cm]. For validation Criterion 1, the mean difference ±SD between the test device and reference BP readings (N = 255) was 1.3 ± 6.5/3.6 ± 5.9 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual (N = 85) was 5.70/5.25 mmHg (systolic/diastolic; threshold ≤6.82/5.89 mmHg).
CONCLUSIONS: The automated oscillometric home BP monitor AVITA BPM82 comfortably fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01 in a general population and can be recommended for clinical use.
METHODS: Participants were recruited to fulfill the age, sex, blood pressure (BP), and limb distribution criteria of the AAMI/ESH/ISO Universal Standard and its Amendment 1.2020-01 in a general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-33 (medium) and 33-42 cm (large).
RESULTS: One-hundred-one individuals were recruited and 85 were analyzed [mean age 57.3 ± 15.1 (SD) years, 46 men, arm circumference 32 ± 5.1 cm, range 22.3-42 cm]. For validation Criterion 1, the mean difference ±SD between the test device and reference BP readings (N = 255) was 1.3 ± 6.5/3.6 ± 5.9 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual (N = 85) was 5.70/5.25 mmHg (systolic/diastolic; threshold ≤6.82/5.89 mmHg).
CONCLUSIONS: The automated oscillometric home BP monitor AVITA BPM82 comfortably fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01 in a general population and can be recommended for clinical use.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app