We have located links that may give you full text access.
Continuous intraoperative neuromonitoring of the recurrent laryngeal nerve during lung operations.
Journal of Thoracic Disease 2023 August 32
BACKGROUND: The recurrent laryngeal nerve (RLN), especially on the left side, is particularly vulnerable during lung operations. Therefore, continuous intraoperative neuromonitoring (cIONM) would be desirable. With the use of a double-lumen tube (DLT) for single-lung ventilation, there is some uncertainty where the recording electrode should be positioned. The aim of this study was to assess the feasibility of this technique and to predict the ideal position of a single recording electrode.
METHODS: Patients scheduled to undergo left thoracotomy due to pulmonary pathologies, two adhesive electrodes were affixed consecutively above the proximal cuff of the DLT prior to intubation for a precise location of the recording from the vocal cords. Following thoracotomy, the vagus nerve alongside the aortic arch was exposed. A continuous stimulation probe (Saxophone® electrode, Dr. Langer Medical, Waldkirch, Germany) was placed around the nerve. The signals of the respective electrode were recorded and analyzed with regard to previously defined limits of positive signaling.
RESULTS: Strong signals with an amplitude of at least 165 µV were recorded in 18 of 20 patients. In these patients 100% of the signals were recorded via the distally located electrode. An additional signal was recorded via the proximal electrode in four patients. cIONM was found to be a safe and reproducible technique.
CONCLUSIONS: The described technique enables the use of cIONM of the RLN throughout the entire course of the surgical procedure.
METHODS: Patients scheduled to undergo left thoracotomy due to pulmonary pathologies, two adhesive electrodes were affixed consecutively above the proximal cuff of the DLT prior to intubation for a precise location of the recording from the vocal cords. Following thoracotomy, the vagus nerve alongside the aortic arch was exposed. A continuous stimulation probe (Saxophone® electrode, Dr. Langer Medical, Waldkirch, Germany) was placed around the nerve. The signals of the respective electrode were recorded and analyzed with regard to previously defined limits of positive signaling.
RESULTS: Strong signals with an amplitude of at least 165 µV were recorded in 18 of 20 patients. In these patients 100% of the signals were recorded via the distally located electrode. An additional signal was recorded via the proximal electrode in four patients. cIONM was found to be a safe and reproducible technique.
CONCLUSIONS: The described technique enables the use of cIONM of the RLN throughout the entire course of the surgical procedure.
Full text links
Trending Papers
Monitoring Macro- and Microcirculation in the Critically Ill: A Narrative Review.Avicenna Journal of Medicine 2023 July
Euglycemic Ketoacidosis in Two Patients Without Diabetes After Introduction of Sodium-Glucose Cotransporter 2 Inhibitor for Heart Failure With Reduced Ejection Fraction.Diabetes Care 2023 November 22
ASA Consensus-based Guidance on Preoperative Management of Patients on Glucagon-like Peptide-1 Receptor Agonists.Anesthesiology 2023 November 21
Tranexamic Acid for Traumatic Injury in the Emergency Setting: A Systematic Review and Bias-Adjusted Meta-Analysis of Randomized Controlled Trials.Annals of Emergency Medicine 2023 November 22
Association between postinduction hypotension and postoperative mortality: a single-centre retrospective cohort study.Canadian Journal of Anaesthesia 2023 November 22
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app