Novel T1 Analysis Method to Address Reduced Measurement Accuracy Due to Irregular Heart Rate Variability in Myocardial T1 Mapping Using Polarity-corrected Inversion Time Preparation.

PURPOSE: Polarity-corrected inversion time preparation (PCTIP), a myocardial T1 mapping technique, is expected to reduce measurement underestimation in the modified Look-Locker inversion recover method. However, measurement precision is reduced, especially for heart rate variability. We devised an analysis using a recurrence formula to overcome this problem and showed that it improved the measurement accuracy, especially at high heart rates. Therefore, this study aimed to determine the effect of this analysis on the accuracy and precision of T1 measurements for irregular heart rate variability.

METHODS: A PCTIP scan using a 3T MRI scanner was performed in phantom experiment. We generated the simulated R-waves required for electrocardiogram (ECG)-gated acquisition using a signal generator set to 30 combinations. T1 map was generated using the signal train of the PCTIP images by nonlinear curve fitting using conventional and recurrence formulas. Accuracy against reference T1 and precision of heart rate variability were evaluated. To evaluate the fitting accuracy of both analyses, the relative fitting error was calculated.

RESULTS: For the longer T1, the fitting error was larger than the short T1, with the conventional analysis showing 10.1±2.0%. The recurrence formula analysis showed a small fitting error less than 1%, which was consistent for all heart rate variability patterns. In the conventional analysis, the accuracy, especially for longer T1, showed a large underestimation of the measurements and poor linearity. However, in the recurrence formula analysis, the accuracy improved at a long T1, and linearity also improved. The Bland-Altman plot showed that it varied greatly depending on the heart rate variability pattern for the longer T1 in the conventional analysis, whereas the recurrence formula analysis suppressed this variation.

CONCLUSION: T1 analysis of PCTIP using the recurrence formula analysis achieved accurate and precise T1 measurements, even for irregular heart rate variability.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app