Comparison of Different Methods of Measuring Finger Range of Motion via Telehealth.

PURPOSE: This study examined the accuracy and reliability of measuring total motion of the fingers via telehealth using the following three different methods: (1) goniometry, (2) visual estimation, and (3) electronic protractor. Measurements were compared with in-person measurement, which was assumed to be the reference standard.

METHODS: Thirty clinicians measured finger range of motion from prerecorded videos of a mannequin hand with articulating fingers, which was posed in extension and flexion that simulated a telehealth visit, using a goniometer with results blinded to the clinician (blinded goniometry), visual estimation, and an electronic protractor, in random order. Total motion was calculated for each finger and for all four fingers in sum. The experience level, familiarity with measuring finger range of motion, and opinions of measurement difficulty were assessed.

RESULTS: Measurement with the electronic protractor was the only method equivalent to the reference standard within 20°. Remote goniometer and visual estimation did not fall within the acceptable error margin of equivalence, and both underestimated total motion. Electronic protractor also had the highest interrater reliability (intraclass correlation [upper limit, lower limit], 0.95 [0.92, 0.95]); goniometry (intraclass correlation, 0.94 [0.91, 0.97]) was nearly identical, whereas visual estimation (intraclass correlation, 0.82 [0.74, 0.89]) was much lower. Clinicians' experience and familiarity with range of motion measurements had no relationship with the findings. Clinicians reported visual estimation as the most difficult (80%) and electronic protractor as the easiest method (73%).

CONCLUSIONS: This study showed that traditional in-person forms of measurement underestimate finger range of motion via telehealth; a new computer-based method (ie, electronic protractor) was found to be more accurate.

CLINICAL RELEVANCE: The use of an electronic protractor can be beneficial to clinicians measuring range of motion in patients virtually.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app