In vivo quantitative MRI: T 1 and T 2 measurements of the human brain at 0.064 T.
Magma 2023 May 21
OBJECTIVE: To measure healthy brain [Formula: see text] and [Formula: see text] relaxation times at 0.064 T.
MATERIALS AND METHODS: [Formula: see text] and [Formula: see text] relaxation times were measured in vivo for 10 healthy volunteers using a 0.064 T magnetic resonance imaging (MRI) system and for 10 test samples on both the MRI and a separate 0.064 T nuclear magnetic resonance (NMR) system. In vivo [Formula: see text] and [Formula: see text] values are reported for white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) for automatic segmentation regions and manual regions of interest (ROIs).
RESULTS: [Formula: see text] sample measurements on the MRI system were within 10% of the NMR measurement for 9 samples, and one sample was within 11%. Eight [Formula: see text] sample MRI measurements were within 25% of the NMR measurement, and the two longest [Formula: see text] samples had more than 25% variation. Automatic segmentations generally resulted in larger [Formula: see text] and [Formula: see text] estimates than manual ROIs.
DISCUSSION: [Formula: see text] and [Formula: see text] times for brain tissue were measured at 0.064 T. Test samples demonstrated accuracy in WM and GM ranges of values but underestimated long [Formula: see text] in the CSF range. This work contributes to measuring quantitative MRI properties of the human body at a range of field strengths.
MATERIALS AND METHODS: [Formula: see text] and [Formula: see text] relaxation times were measured in vivo for 10 healthy volunteers using a 0.064 T magnetic resonance imaging (MRI) system and for 10 test samples on both the MRI and a separate 0.064 T nuclear magnetic resonance (NMR) system. In vivo [Formula: see text] and [Formula: see text] values are reported for white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) for automatic segmentation regions and manual regions of interest (ROIs).
RESULTS: [Formula: see text] sample measurements on the MRI system were within 10% of the NMR measurement for 9 samples, and one sample was within 11%. Eight [Formula: see text] sample MRI measurements were within 25% of the NMR measurement, and the two longest [Formula: see text] samples had more than 25% variation. Automatic segmentations generally resulted in larger [Formula: see text] and [Formula: see text] estimates than manual ROIs.
DISCUSSION: [Formula: see text] and [Formula: see text] times for brain tissue were measured at 0.064 T. Test samples demonstrated accuracy in WM and GM ranges of values but underestimated long [Formula: see text] in the CSF range. This work contributes to measuring quantitative MRI properties of the human body at a range of field strengths.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app