Deviation of DBS Recording Microelectrodes during Insertion Assessed by Intraoperative CT.

INTRODUCTION: Intraoperative microelectrodes recording with the Ben Gun microdrive system are often used during DBS surgery. An accurate location of these microelectrodes will directly influence the interest of this recording. We have studied the imprecision of implantation of these microelectrodes.

METHODS: We have analyzed the stereotactic position of 135 microelectrodes implanted with the Ben Gun microdrive during DBS surgery of 16 patients with advanced Parkinson's disease. An intracranial CT was obtained and integrated to a stereotactic planification system. We recorded the stereotactic coordinates of the 5 microelectrodes inserted simultaneously in a cross-shape. The coordinates of each microelectrode were compared with coordinates of the other 4 electrodes inserted simultaneously with the Ben Gun and visible on the same iCT image. Thus, this procedure avoids errors from image fusion and from brain shift. We calculate (1) the three-dimensional Euclidian deviation of microelectrodes, (2) the deviation in X- and Y-axes on reconstructed probe's eye view MR images, and (3) the deviation from the 2-mm theoretical distance between the central electrode and 4 satellite microelectrodes.

RESULTS: The median deviation was 0.64 mm in 3-D and 0.58 mm in 2-D probe's eye view. Satellite electrodes were located from the central electrode theoretically at 2.0 mm and practically within the range 1.9-2.1 mm, 1.5-2.5 mm, 1.0-3.0 mm, and 0.5-3.5 mm for, respectively, 9.3%, 53.7%, 88.0%, and 98.1%, thus highlighting the significant deviation from the theoretical distance. Position imprecisions were similar for the 4 satellite microelectrodes. The imprecision was similar in X-axis and Y-axes and statistically less in Z-axis. For bilateral implantation, the second implantation of the same patient was not associated with a greater risk of deviation of the microelectrodes than for the first side implanted.

CONCLUSION: A significant percentage of microelectrodes for MER can deviate substantially from their theoretical target during DBS procedures. An iCT can be used to estimate the potential deviation of microelectrodes and improve the interpretation of MER during the procedure.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app