The reversed umbrella: displacement of a Contour device into an MCA aneurysm 18 months after treatment.

SummaryThe Contour is a promising new device designed to treat large-neck intra-cranial aneurysms. We describe a first case of Contour device displacement 18 months after initial treatment.A patient with a 10 mm unruptured right middle cerebral artery bifurcation aneurysm was treated with a 9 mm Contour. The device was correctly positioned at the neck during treatment and at the 6 month angiography follow-up. At 18 months follow-up we noticed a full displacement of the device into the aneurysm dome. The Contour had a reversed shape and the aneurysm was still fully opacified. No neurological event occurred during the whole follow-up. Contour may be a promising tool but needs to be assessed over a long-term period.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app