Validation of Blood Pressure Devices as per 2020 World Health Organization Technical Specifications: A Scoping Review of Global Literature.

Hypertension 2023 March 14
BACKGROUND: To identify and summarize the global research literature on validation of automated noninvasive blood pressure measurement devices (BPMDs) with upper arm cuff, develop a repository of validated BPMDs in compliance with the 2020 World Health Organization technical specifications, and identify challenges and gaps in evidence base on validated BPMDs.

METHODS: A scoping review was conducted. Primary research validating BPMDs complying with the 2020 World Health Organization technical specifications (ie, semiautomated/automated noninvasive devices with upper arm cuff), published in English between January 2000 and December 2021, was included. We searched MEDLINE, Web of Science, Scopus, EMBASE, CINAHL, CENTRAL, ProQuest and the dabl website.

RESULTS: We included 269 studies validating 251 BPMDs across 89 manufacturers. Omron (29%), Microlife (10%), and A&D Company (8%) were the top 3 manufacturers. The 3 most frequently used validation protocols were the European Society of Hypertension-international protocol 2002 (27%), European Society of Hypertension-international protocol 2010 (25%), and modified British Hypertension Society protocol 1993 (16%), respectively. Nearly 45% of the validated BPMDs were intended for use in clinical settings, 38% were for home or self-measurement use, and 48% were for general adults. Most studies reported that BPMDs passed the validation criteria. There was inadequate reporting across studies, especially pertaining to validation settings.

CONCLUSIONS: Most BPMDs fulfilled the validation criteria. However, there are considerable gaps in BPMD research in terms of geographical representation, including specific target populations and diseases/conditions, and a range of arm circumferences. Additionally, a potential strategy is required to accelerate the adoption of the Association for the Advancement of Medical Instrumentation (AAMI)/European Society of Hypertension/International Organization for Standardization Universal Standard (International Organization for Standardization 81060-2:2018) for BPMD validation.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app