The Validation of Cardiac Index and Stroke-Volume Variation Measured by the Pulse-Wave Transit Time-Analysis Versus Conventional Pulse-Contour Analysis After Off-Pump Coronary Artery Bypass Grafting: Observational Study.
Journal of Cardiothoracic and Vascular Anesthesia 2023 Februrary 12
OBJECTIVE: To compare the reliability of cardiac index (CI) and stroke-volume variation (SVV) measured by the pulse-wave transit-time (PWTT) method using estimated continuous cardiac output (esCCO) technique with conventional pulse-contour analysis after off-pump coronary artery bypass grafting (OPCAB).
DESIGN: A single-center, prospective, observational study.
SETTING: At a 1,000-bed university hospital.
PARTICIPANTS: A total of 21 patients were enrolled after elective OPCAB.
INTERVENTIONS: The study authors performed a method comparison study with simultaneous measurement of CI and SVV based on the esCCO technique (CIesCCO and esSVV, correspondingly) and pulse-contour analysis (CIPCA and SVVPCA , correspondingly). As a secondary analysis, they also assessed the trending ability of CIesCCO versus CIPCA. MEASUREMENTS AND MAIN RESULTS: The authors analyzed 178 measurement pairs for CI, and 174 pairs for SVV during the 10 study stages. The mean bias between CIesCCO and CIPCA was 0.06 L min/m2 , with limits of agreement of ± 0.92 L min/m2 and a percentage error (PE) of 35.3%. The analysis of the trending ability of CI measured by PWTT revealed a concordance rate of 70%. The mean bias between esSVV and SVVPCA was -6.1%, with limits of agreement of ± 15.5% and a PE of 137%.
CONCLUSIONS: The overall performance of CIesCCO and esSVV versus CIPCA and SVVPCA is not clinically acceptable. A further improvement of the PWTT algorithm may be required for an accurate and precise assessment of CI and SVV.
DESIGN: A single-center, prospective, observational study.
SETTING: At a 1,000-bed university hospital.
PARTICIPANTS: A total of 21 patients were enrolled after elective OPCAB.
INTERVENTIONS: The study authors performed a method comparison study with simultaneous measurement of CI and SVV based on the esCCO technique (CIesCCO and esSVV, correspondingly) and pulse-contour analysis (CIPCA and SVVPCA , correspondingly). As a secondary analysis, they also assessed the trending ability of CIesCCO versus CIPCA. MEASUREMENTS AND MAIN RESULTS: The authors analyzed 178 measurement pairs for CI, and 174 pairs for SVV during the 10 study stages. The mean bias between CIesCCO and CIPCA was 0.06 L min/m2 , with limits of agreement of ± 0.92 L min/m2 and a percentage error (PE) of 35.3%. The analysis of the trending ability of CI measured by PWTT revealed a concordance rate of 70%. The mean bias between esSVV and SVVPCA was -6.1%, with limits of agreement of ± 15.5% and a PE of 137%.
CONCLUSIONS: The overall performance of CIesCCO and esSVV versus CIPCA and SVVPCA is not clinically acceptable. A further improvement of the PWTT algorithm may be required for an accurate and precise assessment of CI and SVV.
Full text links
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app