Imprecision remains to be improved in the measurement of serum cystatin C with heterogeneous systems.
Clinical Chemistry and Laboratory Medicine : CCLM 2023 March 7
OBJECTIVES: Except for the large bias of some measurement systems for serum cystatin C (CysC) measurements, unacceptable imprecision has been observed for the heterogenous system. This study analyzed the external quality assessment (EQA) results in 2018-2021 to provide an insight into the imprecision of CysC assays.
METHODS: Five EQA samples were sent to participating laboratories every year. Participants were divided into reagent/calibrator-based peer groups, for which the robust mean of each sample and robust coefficient of variation (CV) were calculated by Algorithm A from ISO 13528. Peers with more than 12 participants per year were selected for further analysis. The limit of CV was determined to be 4.85% based on clinical application requirements. The concentration-related effect on CVs was investigated using logarithmic curve fitting; the difference in medians and robust CVs between instrument-based subgroups was also evaluated.
RESULTS: The total number of participating laboratories increased from 845 to 1,695 in four years and heterogeneous systems remained the mainstream (≥85%). Of 18 peers with ≥12 participants, those using homogeneous systems showed relatively steady and small CVs over four years, with the mean four-year CVs ranging from 3.21 to 3.68%. Some peers using heterogenous systems showed reduced CVs over four years, while 7/15 still had unacceptable CVs in 2021 (5.01-8.34%). Six peers showed larger CVs at the low or high concentrations, and some instrument-based subgroups presented greater imprecision than others.
CONCLUSIONS: More efforts should be made to improve the imprecision of heterogeneous systems for CysC measurement.
METHODS: Five EQA samples were sent to participating laboratories every year. Participants were divided into reagent/calibrator-based peer groups, for which the robust mean of each sample and robust coefficient of variation (CV) were calculated by Algorithm A from ISO 13528. Peers with more than 12 participants per year were selected for further analysis. The limit of CV was determined to be 4.85% based on clinical application requirements. The concentration-related effect on CVs was investigated using logarithmic curve fitting; the difference in medians and robust CVs between instrument-based subgroups was also evaluated.
RESULTS: The total number of participating laboratories increased from 845 to 1,695 in four years and heterogeneous systems remained the mainstream (≥85%). Of 18 peers with ≥12 participants, those using homogeneous systems showed relatively steady and small CVs over four years, with the mean four-year CVs ranging from 3.21 to 3.68%. Some peers using heterogenous systems showed reduced CVs over four years, while 7/15 still had unacceptable CVs in 2021 (5.01-8.34%). Six peers showed larger CVs at the low or high concentrations, and some instrument-based subgroups presented greater imprecision than others.
CONCLUSIONS: More efforts should be made to improve the imprecision of heterogeneous systems for CysC measurement.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app