Corneal Higher-Order Aberrations Measurements: Precision of SD-OCT/Placido Topography and Comparison with a Scheimpflug/Placido Topography in Eyes After Small-Incision Lenticule Extraction.

INTRODUCTION: The aim of this study was to evaluate the measurements of corneal higher-order aberrations (HOAs) obtained by a new anterior segment optical coherence tomography (OCT) technique combined with a Placido topographer (the MS-39 device) in eyes with prior small-incision lenticule extraction (SMILE) and compare them to the measurements obtained by a Scheimpflug camera combined with a Placido topographer (the Sirius device).

METHODS: A total of 56 eyes (56 patients) were included in this prospective study. Corneal aberrations were analyzed for the anterior, posterior, and total cornea surfaces. Within-subject standard deviation (Sw ), test-retest repeatability (TRT), and intraclass correlation coefficient (ICC) were used to assess the intraobserver repeatability and interobserver reproducibility. The differences were evaluated by paired t-test. Bland-Altman plots and 95% limits of agreement (95% LoA) were used to evaluate the agreement.

RESULTS: High repeatability was observed for anterior and total corneal parameters, with Sw value < 0.07, TRT ≤ 0.16, and ICCs > 0.893, but not trefoil. For the posterior corneal parameters, ICCs varied from 0.088 to 0.966. Regarding interobserver reproducibility, all Sw values were ≤ 0.04 and TRT ≤ 0.11. ICCs ranged from 0.846 to 0.989, from 0.432 to 0.972, and from 0.798 to 0.985 for the anterior, total, and posterior corneal aberrations parameters, respectively. The mean difference in all aberrations was ≤ 0.05 μm. All parameters showed a narrow 95% LoA.

CONCLUSION: The MS-39 device achieved high precision in both anterior and total corneal measurements; the precision of posterior corneal higher-order RMS, astigmatism II, coma, and trefoil was lower. The two technologies used by the MS-39 and Sirius devices can be used interchangeably for measuring corneal HOAs after SMILE.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app