Single-Center Experience With the Seraph-100® Microbind® Affinity Blood Filter in Patients With SARS-CoV-2 Infection and Septic Shock at a Military Treatment Facility.

Military Medicine 2023 Februrary 29
INTRODUCTION: Coronavirus disease 2019 (COVID-19) infection is characterized by a dysregulated inflammatory response, which may result in severe hemodynamic instability and septic shock. The Seraph-100® Microbind® Affinity Blood Filter is a commercially available extracorporeal pathogen absorbent device with the ability to bind pathogens and cytokines present within the blood. Our study aimed to evaluate the efficacy of the Seraph-100® for patients with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection and septic shock.

MATERIALS AND METHODS: A retrospective review was performed to evaluate the use of the Seraph-100® blood filter at a Single Center tertiary-care facility at Brooke Army Medical Center from 2020 to 2021. Nine patients (11 treatments) were completed with the Seraph-100® blood filter in patients with SARS-CoV-2 infection and septic shock. The Seraph-100® blood filter was used in combination with continuous renal replacement therapy and slow low-efficiency dialysis or directly in parallel with the extracorporeal membrane oxygenation circuit.

RESULTS: Within this cohort, there was a statistically significant improvement in the following clinical parameters comparing values to before and after treatment with the Seraph-100® blood filter: Mean arterial pressure (mmHg) (64.2 ± 2.36 vs. 76.2 ± 2.68; P < .001), heart rate (beats per minute) (128 ± 6.6 vs. 100.3 ± 6.07; P < .001), administered fraction of oxygen (%) (74.4 ± 10.58 vs. 60.3 ± 10.35; P < .001), serum lactate (mmol/L) (6.14 ± 1.25 vs. 2.8 ± 1.14; P = .0095), and pH (7.29 ± 0.03 vs. 7.39 ± 0.04; P = .0203). There were statistically significant improvements in vasopressor requirements: norepinephrine (mcg/min) (36.3 ± 4.74 vs. 8.3 ± 3.92; P < .001), vasopressin (units/min) (0.04 ± 0.004 vs. 0.02 ± 0.003; P < .001), dobutamine (mcg/kg/min) (2.3 ± 1.00 vs. 0.2 ± 0.75; P = .006) and angiotensin II (ng/kg/min) (34.5 ± 6.4 vs. 3.3 ± 5.26; P < .001).

CONCLUSIONS: The use of the Seraph-100® blood filter was associated with statistically and clinically significant improvements in hemodynamic parameters and decreased vasopressor requirements in patients with septic shock and SARS-CoV-2 infection.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app