Beam-Shaping Extended Depth of Focus Intraocular Lens: Optical Assessment With Corneas of Increasing Spherical Aberration.
Journal of Refractive Surgery 2023 Februrary
PURPOSE: To assess the optical quality and halo formation of a beam-shaping extended depth-of-focus (EDOF) intraocular lens (IOL) (AcrySof IQ Vivity; Alcon Laboratories, Inc) with corneas of long-range spherical aberration (SA) such as those resulting from myopic laser ablations.
METHODS: The optical quality of the EDOF IOL and a reference monofocal IOL was evaluated with three corneas (A, B, and C, with SA =+0.135, +0.290, +0.540 µm, respectively, for a 5.15-mm pupil at the IOL plane). The through-focus modulation transfer function area (MTFa) curves were obtained between -5.00 and +2.00 diopters (D) defocus range. The halo was also assessed with the three corneas.
RESULTS: Through-focus MTFa curves for a 4.5-mm IOL pupil showed a slight decrease in the maximum MTFa value provided by the EDOF IOL compared to the monofocal IOL in the three corneal situations (A: 45.9 vs 38.6 units; B: 41.1 vs 33.1 units, and C: 26.9 vs 23.8 units). For the 3.0-mm pupil, the EDOF IOL also had lower maximum MTFa than the monofocal IOL; however, the depth-of-focus increased to -2.20 D. With corneas A and B, the halo induced was of low energy with both IOLs. With cornea C, the EDOF IOL created a much larger and intense halo.
CONCLUSIONS: The EDOF IOL provided a good distance optical performance and an extended range of focus of approximately 2.00 D, with a halo of low intensity when evaluated with a corneal SA similar to the one induced by a low to moderate myopic ablation. With a high myopic ablation, the halo induced would be of considerable size and energy. [ J Refact Surg. 2023;39(2):95-102.] .
METHODS: The optical quality of the EDOF IOL and a reference monofocal IOL was evaluated with three corneas (A, B, and C, with SA =+0.135, +0.290, +0.540 µm, respectively, for a 5.15-mm pupil at the IOL plane). The through-focus modulation transfer function area (MTFa) curves were obtained between -5.00 and +2.00 diopters (D) defocus range. The halo was also assessed with the three corneas.
RESULTS: Through-focus MTFa curves for a 4.5-mm IOL pupil showed a slight decrease in the maximum MTFa value provided by the EDOF IOL compared to the monofocal IOL in the three corneal situations (A: 45.9 vs 38.6 units; B: 41.1 vs 33.1 units, and C: 26.9 vs 23.8 units). For the 3.0-mm pupil, the EDOF IOL also had lower maximum MTFa than the monofocal IOL; however, the depth-of-focus increased to -2.20 D. With corneas A and B, the halo induced was of low energy with both IOLs. With cornea C, the EDOF IOL created a much larger and intense halo.
CONCLUSIONS: The EDOF IOL provided a good distance optical performance and an extended range of focus of approximately 2.00 D, with a halo of low intensity when evaluated with a corneal SA similar to the one induced by a low to moderate myopic ablation. With a high myopic ablation, the halo induced would be of considerable size and energy. [ J Refact Surg. 2023;39(2):95-102.] .
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app