Journal Article
Research Support, N.I.H., Extramural
Add like
Add dislike
Add to saved papers

Evaluating a Novel, Portable, Self-Administrable Device ("Beacon") That Measures Critical Flicker Frequency as a Test for Hepatic Encephalopathy.

INTRODUCTION: We compared critical flicker frequency (CFF) thresholds obtained using a novel portable device "Beacon" with thresholds from the commercially available Lafayette Flicker Fusion System (Lafayette-FFS) in patients with cirrhosis.

METHODS: One hundred fifty-three participants with chronic liver disease underwent CFF testing using Beacon and Lafayette-FFS with a method-of-limits and/or forced-choice protocol.

RESULTS: Beacon demonstrated excellent test-retest reliability (intraclass correlation 0.91-0.97) and good correlation with the Lafayette-FFS values (intraclass correlation 0.77-0.84). Forced-choice CFF were on average 4.1 Hz higher than method-of-limits descending CFFs.

DISCUSSION: Beacon can be self-administered by patients with chronic liver disease and cirrhosis to measure CFF, a validated screening test for minimal hepatic encephalopathy.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app