Identification of Active Pulmonary Tuberculosis Among Patients With Positive Interferon-Gamma Release Assay Results: Value of a Deep Learning-based Computer-aided Detection System in Different Scenarios of Implementation.

PURPOSE: To evaluate the accuracy of a deep learning-based computer-aided detection (CAD) system in identifying active pulmonary tuberculosis on chest radiographs (CRs) of patients with positive interferon-gamma release assay (IGRA) results in different scenarios of clinical implementation.

MATERIALS AND METHODS: We collected the CRs of consecutive patients with positive IGRA results. Findings of active pulmonary tuberculosis on CRs were independently evaluated by the CAD and a thoracic radiologist, followed by interpretation using the CAD. Sensitivity and specificity were evaluated in different scenarios: (a) radiologists' interpretation, (b) radiologists' CAD-assisted interpretation, and (c) CAD-based prescreening (radiologists' interpretation for positive CAD results only). We conducted a reader test to compare the accuracy of the CAD with those of 5 radiologists.

RESULTS: Among 1780 patients (men, 53.8%; median age, 56 y), 44 (2.5%) were diagnosed with active pulmonary tuberculosis. The CAD-assisted interpretation exhibited a higher sensitivity (81.8% vs. 72.7%; P=0.046) but lower specificity than the radiologists' interpretation (84.1% vs. 85.7%; P<0.001). The CAD-based prescreening exhibited a higher specificity than the radiologists' interpretation (88.8% vs. 85.7%; P<0.001) at the same sensitivity, with a workload reduction of 85.2% (1780 to 263). In the reader test, the CAD exhibited a higher sensitivity than radiologists (72.7% vs. 59.5%; P=0.005) at the same specificity (88.0%), and CAD-assisted interpretation significantly improved the sensitivity of radiologists' interpretation (72.3%; P<0.001).

CONCLUSIONS: For identifying active pulmonary tuberculosis among patients with positive IGRA results, deep learning-based CAD can enhance the sensitivity of interpretation. CAD-based prescreening may reduce the radiologists' workload at an improved specificity.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app