Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model.

PURPOSE: Previous work used phantoms to calibrate the nonlinear relationship between the gadolinium contrast concentration and the intensity of the magnetic resonance imaging signal. This work proposes a new nonlinear calibration procedure without phantoms and considers the variation of contrast agent mass minimum combined with the multiple input blood flow system. This also proposes a new single-input method with meaningful variables that is not influenced by reperfusion or noise generated by aliasing. The reperfusion in the lung is usually neglected and is not considered by the indicator dilution method. However, in cases of lung cancer, reperfusion cannot be neglected. A new multiple input method is formulated, and the contribution of the pulmonary artery and bronchial artery to lung perfusion can be considered and evaluated separately.

METHODS: The calibration procedure applies the minimum variation of contrast agent mass in 3 different regions: (1) pulmonary artery, (2) left atrium, and (3) aorta. It was compared with four dimensional computerized tomography with iodine, which has a very high proportional relationship between contrast agent concentration and signal intensity.

RESULTS: Nonlinear calibration was performed without phantoms, and it is in the range of phantom calibration. It successfully separated the contributions of the pulmonary and bronchial arteries. The proposed multiple input method was verified in 6 subjects with lung cancer, and perfusion from the bronchial artery, rich in oxygen, was identified as very high in the cancer region.

CONCLUSIONS: Nonlinear calibration of the contrast agent without phantoms is possible. Separate contributions of the pulmonary artery and aorta can be determined.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app