Comparison of single taper and dual taper versions of the same stem design in total hip arthroplasty for primary osteoarthritis.
Journal of Orthopaedics and Traumatology : Official Journal of the Italian Society of Orthopaedics and Traumatology 2023 Februrary 2
BACKGROUND: In total hip arthroplasty (THA), the outcomes of single taper (ST) and dual taper (DT) versions of the same stem design have been scarcely studied. A registry study comparing ST and DT versions of the same stem design was designed, aiming to assess: (1) the survival rates and the hazard ratios for failure; (2) the survival rates and the hazard ratios for failure using stem-focused endpoints.
MATERIAL AND METHODS: A regional arthroplasty registry was interrogated about stem designs with ST and DT versions in cementless THAs performed for primary osteoarthritis. Only the same cup and ceramic-on-ceramic bearings were included: the DT stems had a titanium-on-titanium modularity. Demographic and implant features were recorded. Survival rates and hazard ratios were evaluated and compared. Stem-focused endpoints were also investigated.
RESULTS: A total of 5359 THAs were included, with three stem designs. The two versions of every stem showed different demographics and implant-related features: ST versions were preferentially implanted in heavier young men. For each stem, the two versions had similar survival rates at 5 years (p = 0.076; p = 0.319; p = 0.616) and similar adjusted hazard ratios for failures (p = 0.084; p = 0.308; p = 0.729). When stem-focused endpoints were adopted, the ST and DT versions of the three stems achieved similar survival rates (p = 0.710; p = 0.784; p = 0.983) and similar adjusted hazard ratios (p = 0.647; p = 0.858; p = 0.787). Three neck breakages occurred (0.0007% of all the modular implants).
CONCLUSIONS: ST and DT versions of the same stem design did not show any differences in terms of survival rates and hazard ratios for failures at 5 years.
LEVEL OF EVIDENCE: IV.
MATERIAL AND METHODS: A regional arthroplasty registry was interrogated about stem designs with ST and DT versions in cementless THAs performed for primary osteoarthritis. Only the same cup and ceramic-on-ceramic bearings were included: the DT stems had a titanium-on-titanium modularity. Demographic and implant features were recorded. Survival rates and hazard ratios were evaluated and compared. Stem-focused endpoints were also investigated.
RESULTS: A total of 5359 THAs were included, with three stem designs. The two versions of every stem showed different demographics and implant-related features: ST versions were preferentially implanted in heavier young men. For each stem, the two versions had similar survival rates at 5 years (p = 0.076; p = 0.319; p = 0.616) and similar adjusted hazard ratios for failures (p = 0.084; p = 0.308; p = 0.729). When stem-focused endpoints were adopted, the ST and DT versions of the three stems achieved similar survival rates (p = 0.710; p = 0.784; p = 0.983) and similar adjusted hazard ratios (p = 0.647; p = 0.858; p = 0.787). Three neck breakages occurred (0.0007% of all the modular implants).
CONCLUSIONS: ST and DT versions of the same stem design did not show any differences in terms of survival rates and hazard ratios for failures at 5 years.
LEVEL OF EVIDENCE: IV.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app