Study of analytical error in lipase assay.
BACKGROUND: Lipase and amylase are the most frequently used biomarker for the diagnosis of pancreatic diseases, especially acute pancreatitis. Lipase has better diagnostic accuracy in comparison to amylase for the analysis of acute pancreatitis. However, lipase assay in random access analyzer is sometimes difficult to perform as it is exposed to different types of samples or reagents which may act as interference.
MATERIALS AND METHODS: In our laboratory, we found the raised values (>500 IU/L) of lipase with normal amylase in some samples. However, the immediate rerun of these samples for lipase only showed normal (<80 IU/L) lipase level. To root out this fallacy, we performed reagent and sample carryover studies.
RESULTS: The cause of the falsely raised value of lipase was revealed by reagent carryover studies. All samples which assayed triglyceride (TGL) followed by lipase immediately after it showed elevated (>500 IU/L) lipase value. This is due to the interference of microbial lipase used in TGL reagents. This was corrected by separating the analysis of lipase and TGL into two different instruments.
CONCLUSION: If interference testing is not done, the laboratories are prone to have an analytical error in reporting and hence lead to diagnostic error. Hence, after analyzer installation, interference testing should be included in the validation protocol.
MATERIALS AND METHODS: In our laboratory, we found the raised values (>500 IU/L) of lipase with normal amylase in some samples. However, the immediate rerun of these samples for lipase only showed normal (<80 IU/L) lipase level. To root out this fallacy, we performed reagent and sample carryover studies.
RESULTS: The cause of the falsely raised value of lipase was revealed by reagent carryover studies. All samples which assayed triglyceride (TGL) followed by lipase immediately after it showed elevated (>500 IU/L) lipase value. This is due to the interference of microbial lipase used in TGL reagents. This was corrected by separating the analysis of lipase and TGL into two different instruments.
CONCLUSION: If interference testing is not done, the laboratories are prone to have an analytical error in reporting and hence lead to diagnostic error. Hence, after analyzer installation, interference testing should be included in the validation protocol.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app