First-Attempt Intubation Success Among Emergency Medicine Trainees by Laryngoscopic Device and Training Year: A National Emergency Airway Registry Study.

STUDY OBJECTIVE: We compare intubation first-attempt success with the direct laryngoscope, hyperangulated video laryngoscope, and standard geometry video laryngoscope among emergency medicine residents at various postgraduate years (PGY) of training.

METHODS: We analyzed prospective data from emergency department (ED) patients enrolled in the National Emergency Airway Registry from January 1, 2016 to December 31, 2018 using mixed-effects logistic regression to assess the association between PGY of training and first-attempt success by the device.

RESULTS: Among 15,204 intubations performed by emergency medicine trainees, first-attempt success for PGY-1, PGY-2, and PGY3+ residents, respectively were: 78.8% (95% CI, 75.0 to 82.2%), 81.3% (79.4 to 83.0), and 83.6% (95% CI, 82.1 to 85.1) for direct laryngoscope; 87.2% (95% CI, 84.2 to 89.7), 90.4% (95% CI, 88.8 to 91.9%), and 91.2% (95% CI, 89.8 to 92.5%) for hyperangulated video laryngoscope; and 88.7% (95% CI, 86.1 to 90.9), 90.2% (95% CI, 88.7 to 91.5%), and 94.6% (95% CI 93.9 to 95.3%) for standard geometry video laryngoscope. Direct laryngoscope first-attempt success improved for PGY-2 (adjusted odds ratio [aOR],1.41; 95% CI, 1.09 to 1.82) and PGY-3+ (aOR, 1.76; 1.36 to 2.27) trainees compared to PGY-1. Hyperangulated video laryngoscope success also improved for PGY-2 (aOR, 1.51; 1.1 to 2.05) and PGY-3+ (aOR, 1.56; 1.15 to 2.13) trainees compared to PGY-1. For the standard geometry video laryngoscope, only PGY-3+ (aOR, 1.72; 1.25 to 2.36) was associated with improved first-attempt success compared to PGY-1.

CONCLUSION: Each laryngoscopy device class was associated with improvement in first-attempt success as training progressed. The video laryngoscope outperformed the direct laryngoscope for all operator groups, and PGY-1 trainees achieved higher first-attempt success using a standard geometry video laryngoscope than PGY-3+ trainees using a direct laryngoscope. These findings support the conjecture that in adult patients, a direct laryngoscope should not be routinely used for the first intubation attempt unless clinical circumstances, such as the presence of a soiled airway, would favor its success. These findings need to be validated with prospective randomized clinical trials.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app