Clinical performance of two new, fully integrated molecular platforms used for HIV-1, HBV and HCV viral load analysis, the NeuMoDx 288 and the Alinity m.
Journal of Clinical Virology 2023 January 3
BACKGROUND: Viral load (VL) determination in patients with human immunodeficiency virus type 1 (HIV-1), hepatitis B virus (HBV) and hepatitis C virus (HCV) is essential for proper patient management and follow-up. New molecular platforms have been developed to fully automate these diagnostic assays.
OBJECTIVE: Evaluation of the clinical performance of HIV-1, HBV and HCV VL assays on the Alinity m (Abbott) and NeuMoDx (Qiagen) molecular platforms.
METHOD: Test panels of the three viruses have been compiled of 100 plasma and/or serum samples per target containing non-detectable, non-quantifiable and quantifiable VLs. All samples were retrospectively tested on the Alinity m and NeuMoDx platforms according to manufacturers' instructions.
RESULTS: A total of 74, 86 and 66 samples with valid results for both platforms were included in the HIV-1, HBV and HCV analysis respectively. Overall qualitative agreement of the assays on both platforms was 78% for HIV-1, 93% for HBV and 100% for HCV. Quantitative agreement (less than 0.5 log difference) was shown to be 68% for HIV-1, 68% for HBV and 94% for HCV.
CONCLUSION: The Alinity m and NeuMoDx HCV assay have a comparable performance. Quantification differences in the HIV-1 assay were mostly apparent in the lower VLs and under-quantification of the NeuMoDx HBV assay was observed.
OBJECTIVE: Evaluation of the clinical performance of HIV-1, HBV and HCV VL assays on the Alinity m (Abbott) and NeuMoDx (Qiagen) molecular platforms.
METHOD: Test panels of the three viruses have been compiled of 100 plasma and/or serum samples per target containing non-detectable, non-quantifiable and quantifiable VLs. All samples were retrospectively tested on the Alinity m and NeuMoDx platforms according to manufacturers' instructions.
RESULTS: A total of 74, 86 and 66 samples with valid results for both platforms were included in the HIV-1, HBV and HCV analysis respectively. Overall qualitative agreement of the assays on both platforms was 78% for HIV-1, 93% for HBV and 100% for HCV. Quantitative agreement (less than 0.5 log difference) was shown to be 68% for HIV-1, 68% for HBV and 94% for HCV.
CONCLUSION: The Alinity m and NeuMoDx HCV assay have a comparable performance. Quantification differences in the HIV-1 assay were mostly apparent in the lower VLs and under-quantification of the NeuMoDx HBV assay was observed.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app