Unsupervised and quantitative intestinal ischemia detection using conditional adversarial network in multimodal optical imaging.

PURPOSE: Intraoperative evaluation of bowel perfusion is currently dependent upon subjective assessment. Thus, quantitative and objective methods of bowel viability in intestinal anastomosis are scarce. To address this clinical need, a conditional adversarial network is used to analyze the data from laser speckle contrast imaging (LSCI) paired with a visible-light camera to identify abnormal tissue perfusion regions.

APPROACH: Our vision platform was based on a dual-modality bench-top imaging system with red-green-blue (RGB) and dye-free LSCI channels. Swine model studies were conducted to collect data on bowel mesenteric vascular structures with normal/abnormal microvascular perfusion to construct the control or experimental group. Subsequently, a deep-learning model based on a conditional generative adversarial network (cGAN) was utilized to perform dual-modality image alignment and learn the distribution of normal datasets for training. Thereafter, abnormal datasets were fed into the predictive model for testing. Ischemic bowel regions could be detected by monitoring the erroneous reconstruction from the latent space. The main advantage is that it is unsupervised and does not require subjective manual annotations. Compared with the conventional qualitative LSCI technique, it provides well-defined segmentation results for different levels of ischemia.

RESULTS: We demonstrated that our model could accurately segment the ischemic intestine images, with a Dice coefficient and accuracy of 90.77% and 93.06%, respectively, in 2560 RGB/LSCI image pairs. The ground truth was labeled by multiple and independent estimations, combining the surgeons' annotations with fastest gradient descent in suspicious areas of vascular images. The total processing time was 0.05 s for an image size of 256 × 256 .

CONCLUSIONS: The proposed cGAN can provide pixel-wise and dye-free quantitative analysis of intestinal perfusion, which is an ideal supplement to the traditional LSCI technique. It has potential to help surgeons increase the accuracy of intraoperative diagnosis and improve clinical outcomes of mesenteric ischemia and other gastrointestinal surgeries.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app