AIM: To investigate the accuracy and acceptability of the FreeStyle Libre Flash continuous glucose monitoring system (FSL-CGM) at alternative sites during free living and under experimental conditions.

MATERIALS AND METHODS: Participants with type 1 diabetes were provided with three FSL-CGM sensors applied to the upper arm, the lower back, and the anterior chest. On day 2 or 3, FSL-CGM sensor glucose was compared with venous glucose following a standard meal, during and after an exercise test, and after skin cooling. Participants completed 14-day use of the sensors with concomitant sensor scanning at all sites and capillary glucose tests. The primary outcome was accuracy between sensor sites of 14-day mean glucose. Clarke's error grids, precision absolute relative deviation, and mean absolute relative deviation were calculated.

RESULTS: In the 20 participants, compared with the arm sensor, the accuracy of the back sensor and the chest sensor was 97.9% and 98%, respectively. Under experimental conditions, the arm sensor was more accurate than that of the back and chest. All the sensors recorded higher glucose concentration than venous samples during exercise. The arm and chest sites were most preferred, with the greatest sensor failures from the back.

CONCLUSIONS: The FSL-CGM is clinically accurate when the sensors are applied to alternate chest or back sites. Greater variability occurs during rapid changes in glucose concentration with all sensor sites compared with venous glucose. Understanding these variabilities allows appropriate use of an economically viable continuous glucose monitor.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app