Dimensional compatibility and limitations of tracheal intubation through supraglottic airway devices: a mannequin-based in vitro study.

BACKGROUND: Flexible bronchoscope-guided tracheal intubation through supraglottic airway devices (SGAs) is a well-established element of difficult intubation algorithms. Success can be limited by dimensional incompatibilities between tracheal tubes (TTs) and SGAs.

METHODS: In this in vitro study, we tested the feasibility of TT passage through SGAs, removal of SGAs over TTs, and the ability to guide the flexible bronchoscope with 13 TT brands (internal diameter, 6.5-8.0 mm) and ten different SGAs (#4 and #5) in an intubation mannequin.

RESULTS: We tested 1,040 combinations of SGAs and TTs. Tracheal tube passage failed in 155 (30%) combinations of the five tested first-generation SGAs (117 [46%] with SGA #4, 38 [15%] with SGA #5) and in three (0.6%) combinations of the five tested second-generation SGAs (two [0.8%] with SGA #4 and one [0.4%] with SGA #5). The reason for failed passage of a TT through a first-generation SGA consistently was a too-narrow SGA connector. Removal of the SGA over the TT in the 882 remaining combinations was impossible for all sizes of reinforced TTs, except the Parker Reinforced TT, and was possible for all non-reinforced TTs. Only one combination with SGA #4 and 84 combinations with SGA #5 were not ideal to adequately guide the flexible bronchoscope.

CONCLUSION: Clinically relevant combinations of adult-size TTs and SGAs can be incompatible, rendering flexible bronchoscope-guided tracheal intubation through an SGA impossible. Additional limitations exist regarding removal of the SGA and maneuverability of the flexible bronchoscope.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app