We have located links that may give you full text access.
Working accuracy of pulse oximetry in COVID-19 patients stepping down from intensive care: a clinical evaluation.
BMJ Open Respiratory Research 2020 December
INTRODUCTION: UK guidelines suggest that pulse oximetry, rather than blood gas sampling, is adequate for monitoring of patients with COVID-19 if CO2 retention is not suspected. However, pulse oximetry has impaired accuracy in certain patient groups, and data are lacking on its accuracy in patients with COVID-19 stepping down from intensive care unit (ICU) to non-ICU settings or being transferred to another ICU.
METHODS: We assessed the bias, precision and limits of agreement using 90 paired SpO2 and SaO2 from 30 patients (3 paired samples per patient). To assess the agreement between pulse oximetry (SpO2 ) and arterial blood gas analysis (SaO2 ) in patients with COVID-19, deemed clinically stable to step down from an ICU to a non-ICU ward, or be transferred to another ICU. This was done to evaluate whether the guidelines were appropriate for our setting.
RESULTS: Mean difference between SaO2 and SpO2 (bias) was 0.4%, with an SD of 2.4 (precision). The limits of agreement between SpO2 and SaO2 were as follows: upper limit of 5.2% (95% CI 6.5% to 4.2%) and lower limit of -4.3% (95% CI -3.4% to -5.7%).
CONCLUSIONS: In our setting, pulse oximetry showed a level of agreement with SaO2 measurement that was slightly suboptimal, although within acceptable levels for Food and Drug Authority approval, in people with COVID-19 judged clinically ready to step down from ICU to a non-ICU ward, or who were being transferred to another hospital's ICU. In such patients, SpO2 should be interpreted with caution. Arterial blood gas assessment of SaO2 may still be clinically indicated.
METHODS: We assessed the bias, precision and limits of agreement using 90 paired SpO2 and SaO2 from 30 patients (3 paired samples per patient). To assess the agreement between pulse oximetry (SpO2 ) and arterial blood gas analysis (SaO2 ) in patients with COVID-19, deemed clinically stable to step down from an ICU to a non-ICU ward, or be transferred to another ICU. This was done to evaluate whether the guidelines were appropriate for our setting.
RESULTS: Mean difference between SaO2 and SpO2 (bias) was 0.4%, with an SD of 2.4 (precision). The limits of agreement between SpO2 and SaO2 were as follows: upper limit of 5.2% (95% CI 6.5% to 4.2%) and lower limit of -4.3% (95% CI -3.4% to -5.7%).
CONCLUSIONS: In our setting, pulse oximetry showed a level of agreement with SaO2 measurement that was slightly suboptimal, although within acceptable levels for Food and Drug Authority approval, in people with COVID-19 judged clinically ready to step down from ICU to a non-ICU ward, or who were being transferred to another hospital's ICU. In such patients, SpO2 should be interpreted with caution. Arterial blood gas assessment of SaO2 may still be clinically indicated.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app