1-Year Prospective Evaluation of Clinical Outcomes and Shocks: The Subcutaneous ICD Post Approval Study.

OBJECTIVES: This study evaluated spontaneous arrhythmias and clinical outcomes in the S-ICD System PAS (Subcutaneous Implantable Cardioverter-Defibrillator Post Approval Study) cohort.

BACKGROUND: The U.S. S-ICD PAS trial patient population more closely resembles transvenous ICD cohorts than earlier studies, which included many patients with little structural heart disease and few comorbidities. Early outcomes and low peri-operative complication rates were demonstrated in the S-ICD PAS cohort, but there are no data detailing spontaneous arrhythmias and clinical outcomes.

METHODS: The S-ICD PAS prospective registry included 1,637 de novo patients from 86 U.S. centers. Descriptive statistics, Kaplan-Meier time to event, and multivariate logistic regression were performed using data out to 365 days.

RESULTS: Patients (68.5% men; mean ejection fraction of 32.0%; 42.9% ischemic; 13.4% on dialysis) underwent implantation for primary (76.6%) or secondary prevention indication. The complication-free rate was 92.5%. The appropriate shock (AS) rate was 5.3%. A total of 395 ventricular tachycardia (VT) or fibrillation (VF) episodes were appropriately sensed, with 131 (33.2%) self-terminating. First and final shock efficacy (up to 5 shocks) for the 127 discrete AS episodes were 91.3% and 100.0%, respectively. Discrete AS episodes included 67 monomorphic VT (MVT) and 60 polymorphic VT (PVT)/VF, with first shock efficacy of 95.2% and 86.7%, respectively. There were 19 storm events in 18 subjects, with 84.2% conversion success. Storm episodes were more likely PVT/VF (98 of 137).

CONCLUSIONS: In the first year after implantation, a predominantly primary prevention population with low ejection fraction demonstrated a high complication-free rate and spontaneous event shock efficacy for MVT and PVT/VF arrhythmias at rapid ventricular rates. (Subcutaneous Implantable Cardioverter-Defibrillator System Post Approval Study [S-ICD PAS; NCT01736618).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app