Subcutaneous implantable cardioverter defibrillator lead displacement.

A 52-year-old gentleman with a subcutaneous implantable defibrillator (S-ICD) implanted in 2015 for primary prevention and a background of hypertrophic cardiomyopathy presented via remote monitoring alert (Boston Scientific Latitude NXT) with an aborted charge episode. The episode showed myopotential noise with baseline wander, and an urgent outpatient follow up was arranged. Upon investigation the S-ICD lead had retracted to the pocket of the generator via a reel mechanism, wrapping the lead around the can. This was likely to be caused by insufficient suturing of the lead collar at the base of xiphisternum. The device had automatically disabled the SmartPass filter, which does not currently have an alert mechanism. This case highlights the rare occurrence of S-ICD lead displacement and components of the SmartPass system which can provide an early warning to lead problems. This article is protected by copyright. All rights reserved.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app