Add like
Add dislike
Add to saved papers

Pitfalls of the S-ICD therapy: experiences from a large tertiary centre.

AIM: The subcutaneous ICD (S-ICD) has evolved to a potential first option for many patients who have to be protected from sudden cardiac death. Many trials have underlined a similar performance regarding its effectiveness in relation to transvenous ICDs and have shown the expected benefits concerning infective endocarditis and lead failure. However, there have also been problems due to the peculiarities of the device, such as oversensing and myopotentials. In this study, we present patients from a large tertiary centre suffering from complications with an S-ICD and propose possible solutions.

METHODS AND RESULTS: All S-ICD patients who experienced complications related to the device (n = 40) of our large-scale single-centre S-ICD registry (n = 351 patients) were included in this study. Baseline characteristics, complications occurring and solutions to these problems were documented over a mean follow-up of 50 months. In most cases (n = 23), patients suffered from oversensing (18 cases with T wave or P wave oversensing, 5 due to myopotentials). Re-programming successfully prevented further oversensing episode in 13/23 patients. In 9 patients, generator or lead-related complications, mostly due to infectious reasons (5/9), occurred. Further problems consisted of ineffective shocks in one patient and need for antibradycardia stimulation in 2 patients and indication for CRT in 2 other patients. In total, the S-ICD had to be extracted in 10 patients. 7 of them received a tv-ICD subsequently, 3 patients refused re-implantation of any ICD. One other patient kept the ICD but had antitachycardic therapy deactivated due to inappropriate shocks for myopotential oversensing.

CONCLUSION: The S-ICD is a valuable option for many patients for the prevention of sudden cardiac death. Nonetheless, certain problems are immanent to the S-ICD (limited re-programming options, size of the generator) and should be addressed in future generations of the S-ICD.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app