Platelet transfusion: The effects of a fluid warmer on platelet function.

Transfusion 2020 October 20
Platelet (PLT) transfusions are an important component of hemostatic resuscitation. The AABB has published several guidelines recommending that PLT units should not be infused through blood warming devices.

STUDY DESIGN AND METHODS: Thirty-one units of hospital blood bank apheresis PLTs were obtained. PLT-rich plasma (PRP) aggregometry and thromboelastography (TEG) were performed on the unit samples before and after the units were infused through a Ranger blood/fluid warming device.

RESULTS: There were no differences in any of the aggregometry results before and after infusion of the PLTs through the blood warmer (all P > .32). There was a significant reduction in the TEG maximum amplitude (MA) of 69.8 ± 7.9 mm before and 66.0 ± 8.8 mm after (P < .001) infusion of the PLTs through the blood warmer and α angle 61.8 ± 9.4° before and 59.3 ± 8.2° after (P = .044) infusion of the PLTs through the blood warmer, although both mean values were within normal range for the TEG and not clinically significant. There were very good correlations of aggregometry and TEG results before and after infusion of the PLTs through the blood warmer device.

CONCLUSION: This study did not demonstrate significant deleterious effect on PLT function from infusing apheresis PLT units through a blood warming device by PRP aggregometry. We did detect a statistically significant-but not clinically significant-reduction in TEG MA and α angle. The prohibition of transfusing PLT units though the Ranger blood warming device is not indicated.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app