Commissioning and clinical implementation of Mobius3D and MobiusFX: Experience on multiple linear accelerators.

Physica Medica : PM 2020 October 16
PURPOSE: To provide practical guidelines for Mobius3D commissioning based on experiences of commissioning/clinical implementation of Mobius3D and MobiusFX as patient-specific quality assurance tools on multiple linear accelerators.

METHODS: The vendor-suggested Mobius3D commissioning procedures, including beam model adjustment and dosimetric leaf gap (DLG) optimization, were performed for 6 MV X-ray beams of six Elekta linear accelerators. For the beam model adjustment, beam data, such as the percentage depth dose, off-axis ratio (OAR), and output factor (OF), were measured using a water phantom and compared to the vendor-provided reference values. DLG optimization was performed to determine an optimal DLG correction factor to minimize the mean difference between Mobius3D-calculated and measured doses for multiple volumetric modulated arc therapy (VMAT) plans. Small-field VMAT plans, in which Mobius3D has dose calculate uncertainties, were initially included in the DLG optimization, but excluded later.

RESULTS: The measured beam data were consistent across the six linear accelerators. Relatively large differences between the reference and measured values were observed for the OAR at large off-axis distances (>5 cm) and for the OF for small fields (<3 × 3 cm2 ). The optimal DLG correction factor was 0.6 ± 0.3 (range: 0.3-1.0) with small-field plans and 0.2 ± 0.2 (0.0-0.5) without them.

CONCLUSIONS: A reasonable agreement was found between the vendor-provided reference and measured beam models. DLG optimization results were dependent on the selection of the VMAT plans, requiring careful attention to the known dose calculation uncertainties of Mobius3D when determining a DLG correction factor.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app