Comparative Study
Journal Article
Multicenter Study
Validation Study
Add like
Add dislike
Add to saved papers

Accuracy and Reliability of the Ankle Brachial Index Measurement Using a Multicuff Oscillometric Device Versus the Doppler Method.

OBJECTIVE: Ankle brachial index (ABI) is widely used for the diagnosis of lower extremity artery disease (LEAD). The purpose of this prospective study was to validate the diagnostic ability and reproducibility of a four cuff automated oscillometric device vs. the Doppler method.

METHODS: Patients with suspected LEAD or asymptomatic individuals at risk because of the presence two or more cardiovascular risk factors were enrolled. For each patient, Doppler and oscillometric ABI measurements were repeated by two observers to address intra- and interobserver reproducibility.

RESULTS: In total, 118 patients were evaluated. The prevalence of Doppler ABI (Dop-ABI) ≤ 0.90 was 45.8%. Taking the Dop-ABI as the reference, the sensitivity, specificity, accuracy, positive and negative predictive values of oscillometric ABI (Osc-ABI) during the first measurement by the first observer were 89.1%, 94.4%, 94.1%, 91.8%, and 92.4%, respectively. The concordance for diagnosing ABI ≤0.90 between methods was excellent (kappa coefficients ranging from 0.80 to 0.88 with different observers). Intra-observer reproducibility assessed by intraclass correlation coefficient (ICC) between methods were 0.94 for observer 1 and 0.96 for observer 2. The intra-observer reproducibility using the same method was also excellent (ICC 0.94, 95% confidence interval [CI] 0.91-0.95) for Dop-ABI and 0.95 (95% CI 0.93-0.97) for Osc-ABI). The ICC for interobserver reproducibility using the same method was 0.95 (95% CI 0.92-0.96) for Dop-ABI and 0.96 (95% CI 0.94-0.97) for Osc-ABI.

CONCLUSION: This study validates the excellent diagnostic performances of a four cuff oscillometric device specifically designed for screening for LEAD. The simple measurement method could therefore be advocated in primary care where fast, easy, and reliable methods are suitable.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app