RESEARCH SUPPORT, N.I.H., EXTRAMURAL
Add like
Add dislike
Add to saved papers

Performance of a multiplexed amplicon-based next-generation sequencing assay for HLA typing.

BACKGROUND: Next-generation sequencing (NGS) has enabled efficient high-resolution typing of human leukocyte antigen (HLA) genes with minimal ambiguity. Most commercially available assays amplify individual or subgroup of HLA genes by long-range PCR followed by library preparation and sequencing. The AllType assay simplifies the workflow by amplifying 11 transplant-relevant HLA genes in one PCR reaction. Here, we report the performance of this unique workflow evaluated using 218 genetically diverse samples.

METHODS: Five whole genes (HLA-A/B/C/DQA1/DPA1) and six near-whole genes (HLA-DRB1/DRB345/DQB1/DPB1; excluding exon 1 and part of intron 1) were amplified in a multiplexed, long-range PCR. Manual library preparation was performed per manufacturer's protocol, followed by template preparation and chip loading on the Ion Chef, and sequencing on the Ion S5 sequencer. Pre-specified rules for quality control and repeat testing were followed; technologists were blinded to the reference results. The concordance between AllType and reference results was determined at 2-field resolution. We also describe the ranges of input DNA and library concentrations, read number per sample and per locus, and key health metrics in relation to typing results.

RESULTS: The concordance rates were 98.6%, 99.8% and 99.9% at the sample (n = 218), genotype (n = 1688), and allele (n = 3376) levels, respectively. Three genotypes were discordant, all of which shared the same G group typing results with the reference. Most ambiguous genotypes (116 out of 144, 80.6%) were due to the lack of exon 1 and intron 1 coverage for HLA-DRB1/DRB345/DQB1/DPB1 genes. A broad range of input DNA concentrations and library concentrations were tolerated. Per sample read numbers were adequate for accurate genotyping. Per locus read numbers showed some inter-lot variations, and a trend toward improved inter-locus balance was observed with later lots of reagents.

CONCLUSION: The AllType assay on the Ion Chef/Ion S5 platform offers a robust and efficient workflow for clinical HLA typing at the 2-field resolution. The multiplex PCR strategy simplifies the laboratory procedure without compromising the typing accuracy.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app