Robotic Surgical Assistant Rehearsal: Combining 3-Dimensional-Printing Technology With Preoperative Stereotactic Planning for Placement of Stereoencephalography Electrodes.
Operative Neurosurgery (Hagerstown, Md.) 2020 August 2
BACKGROUND: The use of frameless stereotactic robotic technology has rapidly expanded since the Food and Drug Administration's approval of the Robotic Surgical Assistant (ROSA) in 2012. Although the use of the ROSA robot has greatly augmented stereotactic placement of intracerebral stereoelectroencephalography (sEEG) for the purposes of epileptogenic focus identification, the preoperative planning stages remain limited to computer software.
OBJECTIVE: To describe the use of a 3-dimensionally (3D)-printed patient model in the preoperative planning of ROSA-assisted depth electrode placement for epilepsy monitoring in a pediatric patient.
METHODS: An anatomically accurate 3D model was created and registered in a preoperative rehearsal session using the ROSA platform. After standard software-based electrode trajectory planning, sEEG electrodes were sequentially placed in the 3D model.
RESULTS: Utilization of the 3D-printed model enabled workflow optimization and increased staff familiarity with the logistics of the robotic technology as it relates to depth electrode placement. The rehearsal maneuvers enabled optimization of patient head positioning as well as identification of physical conflicts between 2 electrodes. This permitted revision of trajectory planning in anticipation of the actual case, thereby improving patient safety and decreasing operative time.
CONCLUSION: Use of a 3D-printed patient model enhanced presurgical positioning and trajectory planning in the placement of stereotactic sEEG electrodes for epilepsy monitoring in a pediatric patient. The ROSA rehearsal decreased operative time and increased efficiency of electrode placement.
OBJECTIVE: To describe the use of a 3-dimensionally (3D)-printed patient model in the preoperative planning of ROSA-assisted depth electrode placement for epilepsy monitoring in a pediatric patient.
METHODS: An anatomically accurate 3D model was created and registered in a preoperative rehearsal session using the ROSA platform. After standard software-based electrode trajectory planning, sEEG electrodes were sequentially placed in the 3D model.
RESULTS: Utilization of the 3D-printed model enabled workflow optimization and increased staff familiarity with the logistics of the robotic technology as it relates to depth electrode placement. The rehearsal maneuvers enabled optimization of patient head positioning as well as identification of physical conflicts between 2 electrodes. This permitted revision of trajectory planning in anticipation of the actual case, thereby improving patient safety and decreasing operative time.
CONCLUSION: Use of a 3D-printed patient model enhanced presurgical positioning and trajectory planning in the placement of stereotactic sEEG electrodes for epilepsy monitoring in a pediatric patient. The ROSA rehearsal decreased operative time and increased efficiency of electrode placement.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app