How Does Light Intensity of the Recording Room Affect the Evaluation of Lens and Corneal Clarity by Scheimpflug Tomography?
Cornea 2019 November 6
PURPOSE: To investigate the effects of the light intensity of the recording room where eye examinations are conducted on the evaluation of lens and corneal clarity.
METHOD: This cross-sectional and comparative study used 63 eyes of 32 participants. Corneal and lenticular optical densitometries, pachymetry, corneal volume, anterior segment volume, and anterior chamber depth measurements were taken with the Scheimpflug imaging system when the lights in the recording room were on (80 lux) and off (3 lux). The results were statistically compared.
RESULTS: The mean lens density when the lights were on was 13.61 ± 2.03, whereas that when the lights were off was 8.48 ± 1.48 (P < 0.001). Similarly, the mean corneal density when the lights were on was 15.87 ± 1.36, whereas that when the lights were off was 11.13 ± 0.90 (P < 0.001). The pachymetry, corneal volume, anterior segment volume, and anterior chamber depth measurements did not differ under the different light conditions.
CONCLUSIONS: The densitometry values increased with an increase in light intensity. This study revealed that the lens and corneal densitometry evaluations were significantly affected by the lighting in the recording room.
METHOD: This cross-sectional and comparative study used 63 eyes of 32 participants. Corneal and lenticular optical densitometries, pachymetry, corneal volume, anterior segment volume, and anterior chamber depth measurements were taken with the Scheimpflug imaging system when the lights in the recording room were on (80 lux) and off (3 lux). The results were statistically compared.
RESULTS: The mean lens density when the lights were on was 13.61 ± 2.03, whereas that when the lights were off was 8.48 ± 1.48 (P < 0.001). Similarly, the mean corneal density when the lights were on was 15.87 ± 1.36, whereas that when the lights were off was 11.13 ± 0.90 (P < 0.001). The pachymetry, corneal volume, anterior segment volume, and anterior chamber depth measurements did not differ under the different light conditions.
CONCLUSIONS: The densitometry values increased with an increase in light intensity. This study revealed that the lens and corneal densitometry evaluations were significantly affected by the lighting in the recording room.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app