Is an analytical dose engine sufficient for intensity modulate proton therapy in lung cancer?

OBJECTIVE: To identify a subgroup of lung cancer plans where the analytical dose calculation (ADC) algorithm may be clinically acceptable compared to Monte Carlo (MC) dose calculation in intensity modulated proton therapy (IMPT).

METHODS: Robust-optimised IMPT plans were generated for twenty patients to a dose of 70 Gy (relative biological effectiveness) in 35 fractions in Raystation. For each case, four plans were generated: three with ADC optimisation using the pencil beam algorithm (PB) followed by a final dose calculation with the following algorithms: PB (PB-PB), MC (PB-MC) and MC normalised to prescription dose (PB-MC scaled). A fourth plan was generated where MC optimisation and final dose calculation was performed (MC-MC). Dose comparison and γ analysis (PB-PB vs PB-MC) at two dose thresholds were performed: 20% (D20) and 99% (D99) with PB-PB plans as reference.

RESULTS: Overestimation of CTVD99 and CTVmean was observed in all PB-MC compared to PB-PB plans (median: 3.7 Gy(RBE) (5%) (range: 2.3 to 6.9 Gy(RBE)) and 1.8 Gy(RBE) (3%) (0.5 to 4.6 Gy(RBE)). PB-MC scaled plans resulted in significantly higher CTVD2 compared to PB-PB (median difference: -4 Gy(RBE) (-6%) (-5.3 to -2.4 Gy(RBE)), p ≤ .001). The overall median γ pass rates (3%-3 mm) at D20 and D99 were 93.2% (range:62.2-97.5%) and 71.3 (15.4-92.0%). On multivariate analysis, presence of mediastinal disease and absence of range shifters were significantly associated with high γ pass rates. Median D20 and D99 pass rates with these predictors were 96.0% (95.3-97.5%) and 85.4% (75.1-92.0%). MC-MC achieved similar target coverage and doses to OAR compared to PB-PB plans.

CONCLUSION: In the presence of mediastinal involvement and absence of range shifters Raystation ADC may be clinically acceptable in lung IMPT. Otherwise, MC algorithm would be recommended to ensure accuracy of treatment plans.

ADVANCES IN KNOWLEDGE: Although MC algorithm is more accurate compared to ADC in lung IMPT, ADC may be clinically acceptable where there is mediastinal involvement and absence of range shifters.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app