Novel Application Software for the Semi-Automated Analysis of Infrared Meibography Images.
Cornea 2019 September 3
PURPOSE: To develop semi-automated application software that quickly analyzes infrared meibography images taken with the CSO Sirius Topographer (CSO, Italy) and to compare them to the manual analysis system on the device (Phoenix software platform).
METHODS: A total of 52 meibography images verified as high quality were used and analyzed through manual and semi-automated meibomian gland (MG) detector software in this study. For the manual method, an experienced researcher circumscribed the MGs by putting dots around grape-like clusters in a predetermined rectangular area, and Phoenix software measured the MG loss area by percentage, which took around 10 to 15 minutes. MG loss was graded from 1 (<25%) to 4 (severe >75%). For the semi-automated method, 2 blind physicians (I and II) determined the area to be masked by putting 5 to 6 dots on the raw images and measured the MG loss area using the newly developed semi-automated MG detector application software in less than 1 minute. Semi-automated measurements were repeated 3 times on different days, and the results were evaluated using paired-sample t test, Bland-Altman, and kappa κ analysis.
RESULTS: The mean MG loss area was 37.24% with the manual analysis and 40.09%, 37.89%, and 40.08% in the first, second, and third runs with the semi-automated analysis (P < 0.05). Manual analysis scores showed a remarkable correlation with the semi-automated analysis performed by 2 operators (r = 0.950 and r = 0.959, respectively) (P < 0.001). According to Bland-Altman analysis, the 95% limits of agreement between manual analysis and semi-automated analysis by operator I were between -10.69% and 5% [concordance correlation coefficient (CCC) = 0.912] and between -9.97% and 4.3% (CCC = 0.923) for operator II. The limit of interoperator agreement in semi-automated analysis was between -4.89% and 4.92% (CCC = 0.973). There was good to very good agreement in grading between manual and semi-automated analysis results (κ 0.76-0.84) and very good interoperator agreement with semi-automated software (κ 0.91) (P < 0.001).
CONCLUSIONS: For the manual analysis of meibography images, around one hundred dots have to be put around grape-like clusters to determine the MGs, which makes the process too long and prone to errors. The newly developed semi-automated software is a highly reproducible, practical, and faster method to analyze infrared meibography images with excellent correlation with the manual analysis.
METHODS: A total of 52 meibography images verified as high quality were used and analyzed through manual and semi-automated meibomian gland (MG) detector software in this study. For the manual method, an experienced researcher circumscribed the MGs by putting dots around grape-like clusters in a predetermined rectangular area, and Phoenix software measured the MG loss area by percentage, which took around 10 to 15 minutes. MG loss was graded from 1 (<25%) to 4 (severe >75%). For the semi-automated method, 2 blind physicians (I and II) determined the area to be masked by putting 5 to 6 dots on the raw images and measured the MG loss area using the newly developed semi-automated MG detector application software in less than 1 minute. Semi-automated measurements were repeated 3 times on different days, and the results were evaluated using paired-sample t test, Bland-Altman, and kappa κ analysis.
RESULTS: The mean MG loss area was 37.24% with the manual analysis and 40.09%, 37.89%, and 40.08% in the first, second, and third runs with the semi-automated analysis (P < 0.05). Manual analysis scores showed a remarkable correlation with the semi-automated analysis performed by 2 operators (r = 0.950 and r = 0.959, respectively) (P < 0.001). According to Bland-Altman analysis, the 95% limits of agreement between manual analysis and semi-automated analysis by operator I were between -10.69% and 5% [concordance correlation coefficient (CCC) = 0.912] and between -9.97% and 4.3% (CCC = 0.923) for operator II. The limit of interoperator agreement in semi-automated analysis was between -4.89% and 4.92% (CCC = 0.973). There was good to very good agreement in grading between manual and semi-automated analysis results (κ 0.76-0.84) and very good interoperator agreement with semi-automated software (κ 0.91) (P < 0.001).
CONCLUSIONS: For the manual analysis of meibography images, around one hundred dots have to be put around grape-like clusters to determine the MGs, which makes the process too long and prone to errors. The newly developed semi-automated software is a highly reproducible, practical, and faster method to analyze infrared meibography images with excellent correlation with the manual analysis.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app