Invisible hemolysis in serum samples interferes in NSE measurement.
Tumori 2019 August 10
INTRODUCTION: The accuracy of serum neuron-specific enolase (NSE) measurements is critical, particularly in neurologic diseases and cancer. NSE measurements are compromised by slight, even invisible, hemolysis, which can produce apparently higher NSE levels, leading to inappropriate clinical decisions. In this article, we describe this issue and propose a solution for avoiding incorrect results.
METHODS: Twenty blood samples from donors with NSE values that were within the reference interval were considered. Experimental hemolysis was induced in vitro to examine the relationship between the degree of hemolysis and the increase in serum NSE. The data were then subjected to statistical analysis.
RESULTS: There was excellent correlation ( r 2 0.953) between the degree of hemolysis and the rise in NSE concentration. Each hemolysis unit (equal to 1 mg/dL of free hemoglobin) corresponded to a mean value of 0.29 ± 0.09 ng/mL NSE that was released from red blood cells.
CONCLUSION: The hemolysis index must be measured in every sample with no evident hemolysis before assaying it for NSE. Moreover, if the degree of hemolysis is between 5 and 30 units, the increase in NSE (from 1.5 to 9.0 ng/mL) must be calculated, and the laboratory results should be appended with comments that suggest the approximate rise in NSE.
METHODS: Twenty blood samples from donors with NSE values that were within the reference interval were considered. Experimental hemolysis was induced in vitro to examine the relationship between the degree of hemolysis and the increase in serum NSE. The data were then subjected to statistical analysis.
RESULTS: There was excellent correlation ( r 2 0.953) between the degree of hemolysis and the rise in NSE concentration. Each hemolysis unit (equal to 1 mg/dL of free hemoglobin) corresponded to a mean value of 0.29 ± 0.09 ng/mL NSE that was released from red blood cells.
CONCLUSION: The hemolysis index must be measured in every sample with no evident hemolysis before assaying it for NSE. Moreover, if the degree of hemolysis is between 5 and 30 units, the increase in NSE (from 1.5 to 9.0 ng/mL) must be calculated, and the laboratory results should be appended with comments that suggest the approximate rise in NSE.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app