Comparison of real world and core laboratory lupus anticoagulant results from the Antiphospholipid Syndrome Alliance for Clinical Trials and International Networking (APS ACTION) clinical database and repository.

BACKGROUND: Variability remains a challenge in lupus anticoagulant (LA) testing.

OBJECTIVE: To validate LA test performance between Antiphospholipid Syndrome Alliance for Clinical Trials and International Networking (APS ACTION) Core laboratories and examine agreement in LA status between Core and local/hospital laboratories contributing patients to this prospective registry.

METHODS: Five Core laboratories used the same reagents, analyzer type, protocols, and characterized samples for LA validation. Non-anticoagulated registry samples were retested at the corresponding regional Core laboratories and anticoagulated samples at a single Core laboratory. Categorical agreement and discrepancies in LA status between Core and local/hospital laboratories were analyzed.

RESULTS: Clotting times for the reference/characterized plasmas used for normalized ratios were similar between Core laboratories (CV <4%); precision and agreement for LA positive/negative plasma were similar (all CV ≤5%) in the four laboratories that completed both parts of the validation exercise; 418 registry samples underwent LA testing. Agreement for LA positive/negative status between Core and local/hospital laboratories was observed in 87% (115/132) non-anticoagulated and 77% (183/237) anticoagulated samples. However, 28.7% (120/418) of samples showed discordance between the Core and local/hospital laboratories or equivocal LA results. Some of the results of the local/hospital laboratories might have been unreliable in 24.7% (41/166) and 23% (58/252) of the total non-anticoagulated and anticoagulated samples, respectively. Equivocal results by the Core laboratory might have also contributed to discordance.

CONCLUSIONS: Laboratories can achieve good agreement in LA performance by use of the same reagents, analyzer type, and protocols. The standardized Core laboratory results underpin accurate interpretation of APS ACTION clinical data.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app