Comparison of Standard and Enhanced Pulse Oximeter Auditory Displays of Oxygen Saturation: A Laboratory Study With Clinician and Nonclinician Participants.

BACKGROUND: When engaged in visually demanding tasks, anesthesiologists depend on the auditory display of the pulse oximeter (PO) to provide information about patients' oxygen saturation (SpO2). Current auditory displays are not always effective at providing SpO2 information. In this laboratory study, clinician and nonclinician participants identified SpO2 parameters using either a standard auditory display or an auditory display enhanced with additional acoustic properties while performing distractor tasks and in the presence of background noise.

METHODS: In a counterbalanced crossover design, specialist or trainee anesthesiologists (n = 25) and nonclinician participants (n = 28) identified SpO2 parameters using standard and enhanced PO auditory displays. Participants performed 2 distractor tasks: (1) arithmetic verification and (2) keyword detection. Simulated background operating room noise played throughout the experiment. Primary outcomes were accuracies to (1) detect transitions to and from an SpO2 target range and (2) identify SpO2 range (target, low, or critical). Secondary outcomes included participants' latency to detect target transitions, accuracy to identify absolute SpO2 values, accuracy and latency of distractor tasks, and subjective judgments about tasks.

RESULTS: Participants were more accurate at detecting target transitions using the enhanced display (87%) than the standard display (57%; odds ratio, 7.3 [95% confidence interval {CI}, 4.4-12.3]; P < .001). Participants were also more accurate at identifying SpO2 range using the enhanced display (86%) than the standard display (76%; odds ratio, 2.7 [95% CI, 1.6-4.6]; P < .001). Secondary outcome analyses indicated that there were no differences in performance between clinicians and nonclinicians for target transition detection accuracy and latency, SpO2 range identification accuracy, or absolute SpO2 value identification.

CONCLUSIONS: The enhanced auditory display supports more accurate detection of target transitions and identification of SpO2 range for both clinicians and nonclinicians. Despite their previous experience using PO auditory displays, clinicians in this laboratory study were no more accurate in any SpO2 outcomes than nonclinician participants.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app