Efficacy of pulsed-xenon ultraviolet light for disinfection of high-touch surfaces in an Ecuadorian hospital.

BACKGROUND: Hospital environment in patient care has been linked on healthcare-associated infections (HAI). No touch disinfection technologies that utilize pulsed xenon ultraviolet light has been recognized to prevent infection in contaminated environments. The purpose of this study was: 1) to evaluate the effectiveness of pulsed-xenon ultraviolet light (PX-UV) disinfection for the reduction of bacteria on environmental surfaces of Hospital General Enrique Garcés, and 2) to evaluate the in-vitro efficacy against multi-drug resistance microorganisms.

METHODS: This was a quality-improvement study looking at cleaning and disinfection of patient areas. During the study, a total of 146 surfaces from 17 rooms were sampled in a secondary 329-bed public medical center. Microbiological samples of high-touch surfaces were taken after terminal manual cleaning and after pulsed xenon ultraviolet disinfection. Cleaning staff were blinded to the study purpose and told clean following their usual protocols. For positive cultures PCR identification for carbapenemase-resistance genes (blaKPC, blaIMP, blaVIM, and blaNDM) were analyzed and confirmed by sequencing. The total number of colony forming units (CFU) were obtained and statistical analyses were conducted using Wilcoxon Rank Sum tests to evaluate the difference in CFU between terminal manual cleaning and after pulsed xenon ultraviolet disinfection.

RESULTS: After manual disinfection of 124 surfaces showed a total of 3569 CFU which dropped to 889 CFU in 80 surfaces after pulsed xenon disinfection (p < 0.001). Overall, the surface and environmental contamination was reduced by 75% after PX-UV compared to manual cleaning and disinfection. There were statistically significant decreases in CFU counts of high touch surfaces in OR 87% (p < 0.001) and patient rooms 76% (p < 0.001). Four rooms presented serine carbapenemases blaKPC, and metallo beta-lactamases blaNDM, blaVIM, blaIMP. confirmed by PCR and sequencing. The in-vitro testing with endemic strains found that after five minutes of pulsed xenon ultraviolet exposure an 8-log reduction was achieved in all cases.

CONCLUSION: This study is one of the first of its kind in an Ecuador Hospital. We found that pulsed-xenon ultraviolet disinfection technology is an efficacious complement to the established manual cleaning protocols and guidelines in the significant reduction of MDRO.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app