MENU ▼
Read by QxMD icon Read
search
OPEN IN READ APP
JOURNAL ARTICLE

Perceptual Limits of Optical See-Through Visors for Augmented Reality Guidance of Manual Tasks

Sara Condino, Marina Carbone, Roberta Piazza, Mauro Ferrari, Vincenzo Ferrari
IEEE Transactions on Bio-medical Engineering 2019 May 6
31059421

The focal length of available optical see-through (OST) head-mounted displays (HMDs) is at least 2 m therefore, during manual tasks, the user eye cannot keep in focus both the virtual and real content at the same time. Another perceptual limitation is related to the vergence-accommodation conflict (VAC), this latter being present in binocular vision only. This paper investigates the effect of incorrect focus cues on the user performance, visual comfort and workload during the execution of augmented reality (AR) guided manual task with one of the most advanced OST HMD, the Microsoft HoloLens.

METHODS: An experimental study was designed to investigate the performance of 20 subjects in a connect-the-dots task, with and without the use of AR. The following tests were planned: AR guided monocular and binocular; Naked-eye monocular and binocular. Each trial was analyzed to evaluate the accuracy in connecting dots; NASA Task Load Index and Likert questionnaires were used to assess the workload and the visual comfort.

RESULTS: No statistically significant differences were found in the workload, and in the perceived comfort between the AR guided binocular and monocular test. User performances were significantly better during the Naked eye tests. No statistically significant differences in performances were found in the monocular and binocular tests. The maximum error in AR tests was 5.9 mm.

CONCLUSION: Even if there is a growing interest in using commercial OST-HMD, for guiding high-precision manual tasks, attention should be paid to the limitations of the available technology not designed for the peripersonal space.

Comments

You need to log in or sign up for an account to be able to comment.

No comments yet, be the first to post one!

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
31059421
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"