Accuracy, repeatability and reproducibility of a handheld three-dimensional facial imaging device: the Vectra H1

Charles Savoldelli, Gauthier Benat, Laurent Castillo, Emmanuel Chamorey, Jean-Christophe Lutz
Journal of Stomatology, Oral and Maxillofacial Surgery 2019 March 25
30922712

INTRODUCTION: Three-dimensional imaging of facial surfaces is a useful tool in different fields of cranio-facial, maxillo-facial and facial aesthetic surgery. Many devices that use several stereoscopic cameras are available but require a dedicated room. In contrast, the Vectra H-1 is a handheld device that can be used regardless of location but requires three consecutive acquisition and might therefore provide less accuracy. The aim of this study was to assess the accuracy, repeatability and reproducibility of the Vectra H1 device to validate its use in daily medical practice.

MATERIAL AND METHODS: A comparative analysis of the Vectra H1 device and a digital calliper was performed based on 23 distances measured among 11 facial landmarks. One operator repeated the procedure six times on a 24-year-old male volunteer to assess repeatability, and six operators performed the measurement procedure on a 22-year-old female volunteer to assess reproducibility. Repeatability, reproducibility and accuracy versus the distances measured were successively assessed by testing the correlations between the distances measured versus the coefficient of variation (CV) calculated for repeatability, reproducibility and accuracy.

RESULTS: The CVs for all distances ranged from 0.34% to 1.53% and decreased linearly when distances measured increased, and this correlation was significant (p = 0.0026) for repeatability. The CVs for all distances ranged from 0.23% to 2.90% and decreased linearly as distances measured increased; there was a significant correlation (p = 0.00045) for reproducibility.

CONCLUSIONS: This study shows that the Vectra H1 provides an accurate linear assessment of clinical parameters and allows the accurate analysis of craniofacial morphology. Furthermore, this device costs less and requires less space than other multi-pod devices.

Comments

You need to log in or sign up for an account to be able to comment.

No comments yet, be the first to post one!

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
30922712
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"