Clinical Evaluation of a New Swept-Source Optical Coherence Biometer That Uses Individual Refractive Indices to Measure Axial Length in Cataract Patients.
Ophthalmic Research 2019 March 20
BACKGROUNDS: Although the OCT biometer using individual refractive index is available, comparisons of measurement value and intraocular lens (IOL) power calculation error with other SS-OCT biometers are not known.
OBJECTIVES: To compare the new SS-OCT biometer ARGOS (OCTB1), which uses individual refractive indices to measure axial length, with the IOLMaster 700 (OCTB2) and OA-2000 (OLCR), which use equivalent refractive index.
METHOD: Six hundred and twenty-two eyes of 622 patients who had been diagnosed with cataract were enrolled in the study. Among the 158 eyes that had undergone cataract surgery, the postoperative refractive error was evaluated using the Haigis formula.
RESULTS: The axial length measured by the OCTB1 showed a proportional bias in comparison with the other two biometers and a fixed bias in eyes with an axial length ≥26 mm. No significant difference was found in the median absolute refractive prediction error (p = 0.3278). However, in eyes with an axial length ≥26 mm, the OCTB1 showed myopic error compared with the other two biometers (p < 0.0001).
CONCLUSIONS: In eyes with long axial length, when the conventional IOL calculation was optimized with the equivalent refractive index-based instrument, we need to consider that IOL calculation using OCTB1 tends to cause slightly myopic refractive prediction error.
OBJECTIVES: To compare the new SS-OCT biometer ARGOS (OCTB1), which uses individual refractive indices to measure axial length, with the IOLMaster 700 (OCTB2) and OA-2000 (OLCR), which use equivalent refractive index.
METHOD: Six hundred and twenty-two eyes of 622 patients who had been diagnosed with cataract were enrolled in the study. Among the 158 eyes that had undergone cataract surgery, the postoperative refractive error was evaluated using the Haigis formula.
RESULTS: The axial length measured by the OCTB1 showed a proportional bias in comparison with the other two biometers and a fixed bias in eyes with an axial length ≥26 mm. No significant difference was found in the median absolute refractive prediction error (p = 0.3278). However, in eyes with an axial length ≥26 mm, the OCTB1 showed myopic error compared with the other two biometers (p < 0.0001).
CONCLUSIONS: In eyes with long axial length, when the conventional IOL calculation was optimized with the equivalent refractive index-based instrument, we need to consider that IOL calculation using OCTB1 tends to cause slightly myopic refractive prediction error.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app