Masimo Perfusion Index Versus Doppler for Tourniquet Effectiveness Monitoring.
Journal of Special Operations Medicine : a Peer Reviewed Journal for SOF Medical Professionals 0 December
BACKGROUND: In addition to a plethysmograph, Masimo pulse oximeters display a Perfusion Index (PI) value. This study investigated the possible usefulness of PI for monitoring limb tourniquet arterial occlusion.
METHODS: Tactical Ratcheting Medical Tourniquets were applied to the thighs of 15 subjects. Tightening ended at one ratchet-tooth advance beyond Doppler- indicated occlusion. The times and pressures of Doppler and PI signal absences and returns were recorded.
RESULTS: Intermittent PI signal error occurred in 149 of 450 runs (PI, 33% versus Doppler, 0%; p < .0001). PI signal loss lagged Doppler-indicated occlusion by 19 ± 15 seconds (mean ± standard deviation, p < .0001). PI Signal Return lagged tourniquet release by 13 ± 7 seconds (Doppler Signal Return took 1 ± 1 seconds following tourniquet release; p < .0001). PI failed to detect early Doppler audible pulse return in 30 of 39 occurrences.
CONCLUSION: The PI available on Masimo pulse oximeters is not appropriate for monitoring limb tourniquet effectiveness.
METHODS: Tactical Ratcheting Medical Tourniquets were applied to the thighs of 15 subjects. Tightening ended at one ratchet-tooth advance beyond Doppler- indicated occlusion. The times and pressures of Doppler and PI signal absences and returns were recorded.
RESULTS: Intermittent PI signal error occurred in 149 of 450 runs (PI, 33% versus Doppler, 0%; p < .0001). PI signal loss lagged Doppler-indicated occlusion by 19 ± 15 seconds (mean ± standard deviation, p < .0001). PI Signal Return lagged tourniquet release by 13 ± 7 seconds (Doppler Signal Return took 1 ± 1 seconds following tourniquet release; p < .0001). PI failed to detect early Doppler audible pulse return in 30 of 39 occurrences.
CONCLUSION: The PI available on Masimo pulse oximeters is not appropriate for monitoring limb tourniquet effectiveness.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app