Case report: Atrioventricular block after transcatheter atrial septal closure using the Figulla® Flex II ASD occluder.

We report a 7-year-old male patient who developed severe atrioventricular block after transcatheter closure of the atrial septal defect with an Occlutech Figulla® Flex II ASD occluder (FSO). He had a small aortic rim and the defect measuring 22.3 mm by balloon sizing. When a 24 mm FSO was deployed, he developed Wenckebach second-degree heart block; however, it recovered to sinus rhythm. Hence, the device was implanted. The rhythm deteriorated to a fixed 2:1 heart block within 7 hr. He underwent surgical retrieval of the device and closure of the defect. Intraoperative findings demonstrated the right atrium disk compressing the triangle of Koch, resulting in a small hematoma. The rhythm recovered completely by 7 days after the surgery. Care must be taken when a relatively large device is deployed in a patient with small rims as even "soft and flexible" device like the FSO can injure the endocardium.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app