BACKGROUND: Transcatheter tricuspid valve-in-valve replacement (TVIV) is an emerging therapy for dysfunctional surgical valves in patients with congenital and acquired TV disease. The present study was performed to establish baseline quantitative data for echocardiographic and invasive parameters obtained pre- and immediately post-TVIV.

METHODS: Patients were drawn from the VIVID Registry. This study included two cohorts. The registry cohort included all patients entered in the VIVID registry through February 2017 who had both echocardiographic and invasively measured gradients across the TV. The focused cohort comprised a subset of patients from a single institution who had both pre- and post-TVIV echocardiogram images reviewed offline by a single investigator. The echocardiographic variables measured were based on published guidelines from the American Society of Echocardiography.

RESULTS: Assessment of paired pre- and/or postimplant echocardiographic and invasive pressure measurements (n = 199) showed reasonable correlation between mean TV gradient measured invasively with cardiac catheterization and noninvasively both pre- and post-TVIV (R = 0.72, P < .001), although there was a bias toward the echocardiographic gradient being higher than the invasively measured gradient and sizable discrepancies were reported in several patients. In the focused cohort (n = 42), the mean TV inflow gradient was 9.3 ± 5.0 mm Hg pre- and 5.6 ± 2.3 mm Hg post-TVIV (P < .001). The TV pressure halftime and TV:left ventricular outflow tract Doppler velocity index were 215 ± 94 msec and 3.4 ± 1.2, respectively, at baseline, and 170 ± 44 msec and 2.4 ± 0.6 post-TVIV. Both the Doppler velocity index and the TV E velocity correlated with the mean TV inflow gradient.

CONCLUSIONS: This study provides benchmark data for the echocardiographic assessment of valve function after TVIV. In this population, the significance of an inflow gradient after TVIV should be interpreted in the clinical context. The appropriate threshold for defining dysfunction may differ from the levels proposed for assessment of native or newly placed surgical valves.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app